30 16800 Series Portable Logic Analyzers Service Guide
3 Testing Performance
2 Solder the pin strip to the SMA board mount connector:
a Solder the leads on the left side of the pin strip to the
center conductor of the SMA connector as shown in
the diagram below.
b Solder the leads on the right side of the pin strip to
the inside of the SMA connector’s frame as shown in
the diagram below. Use a small amount of solder.
3 Attach the second SMA board mount connector:
a Re-heat the solder connection made in the previous
step, and attach the second SMA connector, as shown
in the diagram below. Note that the second SMA
connector is upside- down, compared to the first. Add a
little solder to make a good connection.
b Solder the center conductor of the second SMA
connector to the center conductor of the first SMA
connector and the leads on the left side of the pin
strip.
c Rotate the assembly 180 degrees and solder the two
SMA board mount connector frames together.
4 Check your work:
a Ensure that the following four points have continuity
between them: The two pins on the left side of the pin
solder