11
Network analyzer Specification Characteristic (dBm)
Output power (system maximum)
<1 GHz >1 GHz
8712ET, no attenuator — 10.5 7
8712ET, with attenuator — 9.5 6
8712ES — 7.5 4
8714ET, no attenuator — 5.5 2
8714ET, with attenuator — 4.5 1
8714ES — 3.5 –1
Parameter Specification Supplemental information
Port switching time
1
(sec)
Reflection calibration — 0.7
Transmission enhanced calibration — 0.7
Two-port calibration — 0.7
SelfCal time
2
(sec)
Reflection calibration — 1 (typical)
Transmission enhanced calibration — 3 (typical)
Two-port calibration — 8 (typical)
1. This is the time required to switch to any new port configuration, and requires that the Test Set Cal has been performed for 201 points, the new measurement configuration is
201 points, and no new SelfCal occurs during switching.
2. This is the time required to perform the SelfCal for any single port configuration, assuming that the Test Set Cal was performed for 201 points, the new measurement configu-
ration is 201 points, and the measurement bandwidth is “Med Wide” (4000 Hz). SelfCal times for other settings can be found in Table 4-1 of the Agilent 87050E/87075C Users
and Service Guide (87050-90026).