Agilent 85032B/E Type-N calibration kit
1. These specifications and supplementals are valid for reflection (one-port) and transmission (enhanced-response) Test Set Cals, and apply under the following conditions: mea-
surement uses the “Fine” (15 Hz) bandwidth, no averaging, and isolation cal “on”; Test Set Cals use the “Fine” (15 Hz) bandwidth, 16 averages, and assume an isolation calibra-
tion has been performed; the test set must be used with an 8712ET, 8712ES, 8714ET, or 8714ES network analyzer with firmware revision E.06.00 or later; the test set and the
analyzer must have had their performance verified within the last year; both instruments must have warmed up for at least 30 minutes after turn-on; measurements are made at
an environmental temperature of 25 °C ± 5 °C and within ± 1 °C of the last valid Test Set Cal.
2. Corrected system performance is changed to typical when the current temperature has drifted beyond ± 1 °C of the last valid Test Set Cal. Typicals are valid only when the cur-
rent temperature is within 25 °C ± 5 °C, and within ± 1 °C of the most recent SelfCal.
System performance
Transmission/reflection (T/R) calibration (user)
Specification
1
( dB) Typical
2
(dB)
Parameter 3 MHz 1.3 GHz 3 MHz 1.3 GHz 2.2 GHz
to 1.3 GHz to 2.2 GHz to 1.3 GHz to 2.2 GHz to 3 GHz
Directivity 47 40 — — 40
Source match 37 32 — — 30
Load match (reflection calibration) 18 18 23 24 20
Load match (transmission calibration) 14 14 18 18 14
Reflection tracking ±0.1 ±0.1 — — ±0.2
Transmission tracking ±0.1 ±0.1 — — ±0.2