Specification Typical
System bandwidth
Fine Fine Med Wide Wide
15 Hz 15 Hz 4000 Hz 6500 Hz
Network analyzer 3 MHz 1.3 GHz 3 MHz 1.3 GHz 3 MHz 1.3 GHz 3 MHz 1.3 GHz
to to to to to to to to
1.3 GHz 2.2 GHz 1.3 GHz 2.2 GHz 1.3 GHz 2.2 GHz 1.3 GHz 2.2 GHz
System noise floor
1
(dBm)
8712ET/8714ET –97 –95 –106 –103 –82 –79 –42 –39
8712ES/8714ES –88 –86 –100 –97 –75 –72 –42 –39
System dynamic range
2
(dB)
8712ET, no attenuator 101 97 112 108 87 83 49 45
8712ET, with attenuator 100 96 111 107 86 82 48 44
8712ES 89 85 103 99 78 74 45 41
8714ET, no attenuator 99 95 109 105 84 80 46 42
8714ET, with attenuator 97 93 108 104 83 79 45 41
8714ES 86 82 100 96 81 77 42 38
1. Noise floor is defined as the rms value of the trace (in linear format) for a transmission measurement in CW mode, using the “Fine” bandwidth (15 Hz), the test ports
terminated in loads, 0 dBm at the test set source port, and no averaging. This measurement ignores the effects of crosstalk. This is a system specification–the test set
increases the network analyzer noise floor by adding loss to the network analyzer measurement.
2. The system dynamic range is calculated as the difference between the receiver minimum input (noise floor plus calibrated crosstalk) and the system’s maximum output
power. System dynamic range applies to transmission measurements only, since reflection measurements are limited by directivity.