N4962A System Overview
22
Serial BERT 12.5 Gb/s User Guide
2.3 Connector Care
Refer to the N4960-90030 N495xA through N498xA Connector Care Reference
Guide at www.agilent.com/find/N4962A
.
Inspect the connectors for the following:
• Worn or damaged threads
• Scratches to mating surface
• Burrs and loose metal particles
• Dust or foreign material in the space surrounding the center pin (type K
only)
• Ensure that female contacts are straight and aligned
Clean the connectors as described in the following procedure. Cleaning
connectors with alcohol shall only be done with the instruments power cord
removed, and in a well-ventilated area. Allow all residual alcohol moisture to
evaporate, and the fumes to dissipate prior to energizing the instrument.
1. Remove any dust or loose particles using a low-pressure air source.
2. Moisten a lint-free swab with isopropyl alcohol. Do not saturate the
swab.
3. Minimize the wicking of the alcohol into the connector structure.
4. Clean the mating plane surfaces and threads.
5. Allow alcohol to evaporate, and then use a low-pressure air source to
blow surfaces clean.
6. Make sure no particles or residue remains.
7. Inspect connector for damage.