Operation
42
Serial BERT 12.5 Gb/s User Guide
Text
Type
Description
Default
sequence. When the source is used with the
error detector, this value should be set to 1. This
helps with a condition when all zeros are loaded
into the error detector’s input, which can
erroneously cause bad error rate numbers. This
shoul
d be set to 0 when using a different
manufacture
r’s pattern generator, or a known
good signal is used with the N4962A source.
TD= 0010
o Min 0010 (1 µs), max 6500 (650 µs)
The TD (“Training Delay”) option delays the start
of the error detector after the BE
R measurement
has been initiated.
This feature is used in
conjunction with the NoData function. This delay,
listed in 1/10
th
microseconds (default 1us),
allows the PRBS pattern time to transition
through the DUT before measurement start. This
can be very important for DUTs containing a long
electrical length or an optical loop.
APT=0010
o Min 0001 (0.1 ms), max 5000 (500
ms)
The APT (“Autophase Time”) option specifies the
autophase search algorithm measurement time at
each 2
-degree phase point. This option
allows the
user to specify the measurement depth of the
autophase algorithm.
APT allows a tradeoff between the search time
and positioning accuracy of the autophase
algorithm. Because 180 phase points are
measured, this option may increase th
e
autophase
time to a total of 90+ seconds.
The confidence interval of a BER measurement,
with frequency (GHz), gate time (seconds), and
Zobrazit stránku 41