Remote GPIB Interface
72
Serial BERT 12.5 Gb/s User Guide
Command
:SYSTem:STATus? {BITERROR | NODATA | NORXCLK}
Type
Query
Description
Returns the state of the output indicator:
BITERROR
Indicates the current status of the BERT error light; the light is
lit when errors are counted; the query returns 0 when the light
is lit
Indicates the current status of the NoData flag determined by
the instrument (only if NODATA is turned on, see above); the
query returns 0 when the NoData condition is met
NORXCLK
Indicates the current status of the NoRXClk flag; the query
returns 0 when no clock input is detected at the RX CKI port
Results
{0 | 1}
Example
None
Command
:SENSe:APTime?
Type
Query
Description
Returns the current autophase measurement time setting; how long to
measure BER at each of the 180 phase points when performing an autophase
alignment, in 1/10ths of a millisecond.
Results
{1 <= value <= 5000}
Example
See below
Komentáře k této Příručce