Agilent Technologies 8935 Series Uživatelský manuál Strana 161

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 329
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 160
Chapter 6 161
Advanced Operations
Status Reporting
Accessing the Hardware Status Register #1 Group’s Registers
The following sections show the syntax and give programming
examples (using the IBASIC programming language) for the STATus
commands used to access the Hardware Status Register #1 Group’s
registers.
Reading the Condition Register
Example 6-81 Syntax
STATus:HARDware1:CONDition?
Example 6-82 Example
OUTPUT 714;"STAT:HARD1:COND?"
ENTER 714;Register_value
Reading the Transition Filters
Example 6-83 Syntax
STATus:HARDware1:PTRansition?
STATus:HARDware1:NTRansition?
Example 6-84 Example
OUTPUT 714;"STAT:HARD1:PTR?"
ENTER 714;Register_value
Writing the Transition Filters
Example 6-85 Syntax
STATus:HARDware1:PTRansition <integer>
STATus:HARDware1:NTRansition <integer>
3 8 Overpower Protection
Tripped
24 Unused in Test Set
12 External Mike Keyed
0 1 External Battery Voltage Low
Table 6-8 Hardware Status Register Group #1, Condition Register Bit
Assignments
Bit
Number
Binary
Weighting
Condition Comment
Zobrazit stránku 160
1 2 ... 156 157 158 159 160 161 162 163 164 165 166 ... 328 329

Komentáře k této Příručce

Žádné komentáře