HP-IB Command Guidelines
Guidelines for Operation
Guidelines for Attribute Units
When setting the value of measurement functions REFerence,
METer, HLIMit and LLIMit through HP-IB, a non–Attribute Unit
unit-of-measure must be specified in the command string, otherwise
the current Attribute Unit is assumed by the Test Set.
For example, if the Test Set is in a RESET condition and the
command MEAS:AFR:DIST:REF:VAL 10 is sent through HP-IB, the
Test Set will interpret the data as 10%, since % is the RESET
Attribute Unit for the Distortion measurement. Sending the
command, MEAS:AFR:DIST:REF:VAL 10 DBM, would set the
REFerence SET value to 10 dB.
When querying measurement functions REFerence, METer, HLIMit
and LLIMit through HP-IB, the Test Set always returns numeric
values in Attribute Units, regardless of the current Display Units or
HP-IB Units settings. Numeric values are expressed in scientific
notation.
For example, if the REF SET measurement function is displayed as
25% on the Test Set, the value returned through HP-IB is
+2.50000000E+001 (2.5 ¥ 10
1
). Converting the returned value to a
format other than scientific notation must be done
programmatically.
Before changing the Attribute Units for a selected measurement, the
Test Set verifies that all Data Function values can be properly
converted from the current unit-of-measure to the new
unit-of-measure. If it is not possible to properly convert all the
values to the new unit-of-measure, the Attribute Units are not
changed and the following error is generated: HP-IB Error: HP-IB
Units cause invalid conversion of attr.