306 Chapter 9
O:\Manuals\E6380A_Progguid\Book\Errormsg.fm
Error Messages
Text Only HP-IB Errors
In order to save the current Test Set state into a Save/Recall register
more non-volatile RAM will have to be made available. This can be done
by,
reducing the size of the IBASIC program
deleting one or more existing Save/Recall registers
recovering RAM space
The ROM utility RAM_USAGE will display the total amount of
non-volatile RAM installed in the Test Set, the RAM allocation, the
Save/Recall register allocation and the amount of non-volatile RAM
available to IBASIC.
HP-IB Error: Unknown Save/Recall error.
This error can occur on a Test Set when trying to SAVE the instrument
state to a mass storage device with a LIF formatted media. The default
file system for the Test Set is DOS. Refer to “Default File System” on
page 199 for more details.
HP-IB Error: HP-IB Units cause invalid conversion of attr.
This error is generated when trying to change Attribute Units and one
of the Data Function values is set to zero. If this error is encountered
the programmer must change the Data Function settings to values that
can be converted to the new units_of_measure. Refer to Attribute Units
(AUNits)” on page 75 for more details.