150 1100 Series Capillary LC System Manual
6Options
Installing the Micro Column Switching Valve
When required For first time installation or after it was removed
Tools re quired Screwdriver Pozidriv 1 PT3
Wrench 5.5 mm for capillary connections
Preparations for this procedure are:
• The column compartment is open as
described in “Removing the Micro
Column Switching Valve" on page 147
1 If no column switching valve was installed, remove
the RFI-shield and the plastic cover (no longer
used).
2 Replace the valve into its location.
Note
Ensure that during the next steps the flexible
cables close to the heat exchanger assemblies are
not damaged.