
1100 Series Capillary LC System Manual 31
Optimizing Performance 2
• To achieve the best flow stability, especially in the micro mode, %Ripple must
be within acceptable values, typically no worse than 2%.
Fused Silica Capillary issues
• When you connect a capillary (especially at the column) press it smoothly
into the fitting to avoid air gaps. Incorrect setting will result in dispersion
causing tailing or footing peaks.
• Be careful when you bend a Fused Silica Capillary. The diameter must not
be smaller than 40 mm.
• When you replace a part, especially a capillary, clean it with Acetone.
• If a fused silica capillary leaks, do not retighten under flow. Set column
flow to zero, reinsert the capillary, tighten and set new column flow.
• Avoid the use of alkaline solutions (pH > 8.5) which can attack the fused
silica from the capillaries.
• Be careful not to crush capillaries when applying module doors.
• A broken capillary can release silica particles into the system (e.g. cell)
causing problems in the system down-stream of the break.
• A blocked capillary can often be unblocked by back-flushing the capillary.
Acetone is a recommended solvent for capillary back-flushing.
Do not overtighten the fused silica capillaries. see Chapter 3, “Capillaries and Fittings,”
starting on page 45 for information on installing and handling capillaries.
40 mm
Komentáře k této Příručce