1100 Series Capillary LC System Manual 63
Basic System Troubleshooting 4
Wandering Detector Baseline
Critical Decision Determine if the problem is in the DAD, or coming from the LC system:
Remove the flow cell from the DAD. Close the cell cover, and see if the baseline
performance improves.
1 If baseline performance has not improved:
a Replace the lamp(s).
b Evaluate the environment for excessive drafts, temperature changes, etc.
2 If baseline performance has improved, focus attention on the possible
causes and suggested actions below.
Wandering Detector Baseline Possible Causes and Suggested Actions
Possible Causes Suggested Actions Notes
Dirty or defective flow cell Clean or replace the flow cell
The analytical column Bypass the column. Pump directly into the flow
cell. If performance improves, try a new column.
Running with the pump mixer
removed
Reinstall the mixer, and evaluate the
performance of the baseline. If the baseline
performance improves, a solution must be found
which is a compromise between mixing volume
and other chromatographic requirements.
The mixer may have been removed in
an attempt to reduce gradient delay
volume.
"Mixing Noise" when
pumping a binary mobile
phase from two channels.
Try premixing the mobile phase in one bottle,
and pumping 100% from that one solvent
channel.
If baseline performance improves, a solution
must be found which is a compromise between
mixing volume and other chromatographic
requirements. For an isocratic analysis,
premixing and pumping 100% one channel is the
best solution.
This problem occurs when one (or
both) of the solvents has a high
background absorbance at the
detection wavelength. In this case,
the pump’s mixing efficiency may
not be good enough to produce a
sufficiently homogeneous mobile
phase.
The detector reacts to gluts of the
more detectable parts of the solvent
mixture, and baseline disturbances
result.