1100 Series Capillary LC System Manual 171
Performance Specifications 7
Performance Specifications Agilent 1100 Thermostatted Column
Compartment.
All specifications in Table 91 are valid for distilled water at ambient
temperature (25 °C), set point at 40 °C and a flow range from 0.2–5 ml/min.
For flow rates below 100 µl/min the column bracket must be installed
Tabl e 91 Performance Specifications Agilent 1100 Series Thermostatted Column
Compartment
Type Specification
Temperature range 10 degrees below ambient to 80 °C
Temperature stability ± 0.15 °C
Column capacity Three 25 cm -
NOTE: With fused silica capillaries connected, length
limited by bend radii of capillary
Warm-up/cool-down time 5 minutes from ambient to 40 °C
10 minutes from 40 – 20 °C
Internal volume 3 µl left heat exchanger
6 µl right heat exchanger
Communications Controller-area network (CAN), GPIB, RS-232C, APG
Remote: ready, start, stop and shut-down signals, LAN
optional
Safety and maintenance Extensive diagnostics, error detection and display
(through control module and Agilent ChemStation), leak
detection, safe leak handling, leak output signal for
shutdown of pumping system. Low voltages in major
maintenance areas.
GLP features Column-identification module for GLP documentation of
column type, see “Column-Identification System”
Housing All materials recyclable.