1100 Series Capillary LC System Manual 37
Optimizing Performance 2
When to use Alternative Seals
The standard seals for the capillary pump can be used for most applications.
However that use normal phase solvents (for example hexane) are not suitable
for the standard seals and require a different seal when used for a longer time
in the capillary pump. In this case we recommend the use of polypropylene
seals, part number 0905-1420 (pack of 2). These seals have less abrasion
compared to the standard seals.
CAUTION
Polyethylene seals have a limited pressure range 0-200 bar. When used above 200 bar
their lifetime will be significantly reduced. DO NOT apply the seal wear in procedures
performed with the standard seals at 400 bar.