1100 Series DAD and MWD Reference Manual 137
Repairing the Detector 4
6
The assembly may be disassembled using a pair of pliers
to remove the retainer ring. The window may be cleaned or
the complete assembly must be exchanged.
7 Insert the rebuilt or new assembly and slide it in until it is
stopped by the filter lever.
8 Leave slight clearance for movement of the filter lever
between both, the lens assembly and the filter lever, and
fix the setscrew.
Next steps:
Perform a filter test to check the proper function.
Replace the flow cell cover, see “Cleaning or
Exchanging the Holmium Oxide Filter” on page 127.
Replace both lamps.
Replace the front cover.
Retainer ring
Spectro window
Cell support
Washer