1100 Series DAD and MWD Reference Manual 163
Repairing the Detector 4
2 Pull the filter motor off the shaft. Keep the spring in a safe
place.
3 Before installing the filter motor, press the filter lever in
from the other side.
4 While counter-holding the filter lever, press the filter
motor onto the filter shaft until it clicks into the final
position.
Next steps:
• Check that the filter lever is moveable.
• Reinstall the optical unit, see “Installing the Optical Unit
(G1315A/G1365A)” on page 171 or “Removing the
Optical Unit (G1315B/G1365B)” on page 156.
• Reconnect the filter motor cable to connector J18
(G1315/65A) or J19 (G1315/65B) on the detector main
board.
• Reinstall the front cover, top cover and top foam section,
see “Replacing the Foam and the Top Cover” on
page 175.
• Replace the detector into the stack.
• Reconnect the power cable and turn on detector.