Agilent Technologies 1100 Series Servisní příručka Strana 47

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 318
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 46
1100 Series DAD and MWD Reference Manual 47
How to optmize the Detector 2
In a four-component mixture, only biphenyl was recorded. The other three
peaks were suppressed because they did not meet the ratio-qualifier criterion
and therefore the output was set to zero. The characteristic wavelengths
249 nm (λ
1
) and 224 nm (λ
2
) were found from the spectra shown in Figure 17.
Figure 17 Wavelength Selection for Ratio Qualifiers
Figure 18 Selectivity by Ratio Qualifiers
Select 2 characteristic
wavelengths
Signal = WL1/WL2 20%
o-Terphenyl
Biphenyl
Wavelength (nm)
WL1
WL2
Scaled
Signals at 250 nm
No selectivity
With ratio qualifier
250/222 nm = 3.5 20%
Biphenyl
o-Terphenyl
Time (min)
Zobrazit stránku 46
1 2 ... 42 43 44 45 46 47 48 49 50 51 52 ... 317 318

Komentáře k této Příručce

Žádné komentáře