Agilent Technologies 1100 Series Servisní příručka Strana 142

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 318
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 141
142 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
6
Disconnect the lamps from their connectors. 7 Remove the two lamp connectors towards the back by
depressing their connector tabs with a screwdriver.
8 While removing the top foam section completely, route the
lamp cables through the hole in the foam.
Note:
Do not connect a power plug to the detector module
after removing the top covers.
A safety light switch on the main board will turn off
fan (immediately) and electronics (after 30 s) to
avoid the operation with removed covers. An error
will be generated (status lamp will light red) and the
logbook will show an error message.
Zobrazit stránku 141
1 2 ... 137 138 139 140 141 142 143 144 145 146 147 ... 317 318

Komentáře k této Příručce

Žádné komentáře