1100 Series DAD and MWD Reference Manual 179
Repairing the Detector 4
Replacing the Interface Board
When required For all repairs inside the detector or for installation of the board
Part required Interface board (BCD) G1351-68701 with external contacts and BCD outputs,
see “BCD Board” on page 253
Interface board (LAN) see “LAN Communication Interface Board” on
page 255.
Tools required None
1 Install the ESD strap. Move the power lock across the
power inlet.
2 If required, unscrew and remove the interface board. Place
the board on the ESD kit.
3 If required, insert the interface board and fix the screws.
Next steps:
Remove the ESD strap.
Reinstall the module into the stack.
ESD strap
power lock