4 Repairing the Detector
9 Insert the power supply into its location and fix it with the
screws at the rear panel.
10 Press down and clip in the power switch light pipe into the
power supply.
11 Reinstall bottom foam part. Slide it in underneath the leak
drain.
Next steps:
Reinstall the processor board, see “Exchanging the
Detector Main Board” on page 143.
Reinstall the fan assembly, see “Exchanging the Fan
(G1315A/G1365A)” on page 148 or “Replacing Fan Parts
(G1315B/G1365B)” on page 150.
Reinstall the optical unit, see “Installing the Optical Unit
(G1315A/G1365A)” on page 171 or “Removing the
Optical Unit (G1315B/G1365B)” on page 156.
Reinstall the front cover, top cover and top foam section,
see “Replacing the Foam and the Top Cover” on
page 175.
Replace detector into the stack.
Reconnect the power cable and turn on the detector.
Power