172 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
2 Carefully insert the optical unit into the instrument holding
the one hand at back of the heat sink.
3 Reconnect the SCI cable (J23) and the holmium oxide filter
motor (J19) to the detector main board.
4 Replace the two rubber shock absorbers and reconnect
the ground connector.
Next steps:
Reinstall the fan assembly (J16). Ensure the correct
direction of air flow.
Reinstall the flow cell cover, see “Cleaning or
Exchanging the Holmium Oxide Filter” on page 127.
Reinstall the flow cell, see “Exchanging a Flow Cell” on
page 105.
Reinstall the lamps, see “Exchanging a Lamp” on
page 102.
Reinstall the front cover, top cover and top foam section,
see “Replacing the Foam and the Top Cover” on
page 175.
Replace detector into the stack.
Reconnect the power cable and turn on the detector.
NOTE
If a new optical is installed, perform a “Wavelength Recalibration” on page 84 to add the
recalibration parameters into the detector main board’s memory after turning on the
detector.