162 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Replacing the Holmium Oxide Filter Motor
When required If defective
Tools required Screwdriver POZI 1 PT3
hexagonal key 3 mm
Parts required Filter motor assembly G1315-68700 (includes filter lever G1315-45001 and
spring 1460-1510)
For others see “Holmium Oxide Filter” on page 197
NOTE
If the filter motor was removed, the filter lever should not be reused. Always use a new
filter lever to assure correct fit on the filter motor shaft.
Preparations for this procedure:
Turn off the lamp.
Switch off the detector, and disconnect the power cable.
Remove detector from stack and place it on working
bench.
Remove front cover, top cover and top foam section, see
“Removing the Top Cover and Foam” on page 140.
Remove optical unit, see “Removing the Optical Unit
(G1315A/G1365A)” on page 154 or “Removing the
Optical Unit (G1315B/G1365B)” on page 156.
1 Place the optical unit on your bench.