Agilent Technologies InfiniiVision 3000 DSO-X 3054A Servisní příručka Strana 191

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 970
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 190
:ACQuire Commands 7
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 191
:ACQuire:MODE
(see page 860)
Command Syntax
:ACQuire:MODE <mode>
<mode> ::= {RTIMe | SEGMented}
The :ACQuire:MODE command sets the acquisition mode of the
oscilloscope.
The :ACQuire:MODE RTIMe command sets the oscilloscope in real time
mode.
The :ACQuire:MODE SEGMented command sets the oscilloscope in
segmented memory mode.
Query Syntax
:ACQuire:MODE?
The :ACQuire:MODE? query returns the acquisition mode of the
oscilloscope.
Return Format
<mode_argument><NL>
<mode_argument> ::= {RTIM | SEGM}
See Also "Introduction to :ACQuire Commands" on page 187
":ACQuire:TYPE" on page 199
NOTE
The obsolete command ACQuire:TYPE:REALtime is functionally equivalent to sending
ACQuire:MODE RTIMe; TYPE NORMal.