Agilent Technologies InfiniiVision 3000 DSO-X 3054A Servisní příručka Strana 539

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 970
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 538
:SBUS<n> Commands 24
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 539
:SBUS<n>:UART:BAUDrate
(see page 860)
Command Syntax
:SBUS<n>:UART:BAUDrate <baudrate>
<baudrate> ::= integer from 1200 to 8000000 in 100 b/s increments
The :SBUS<n>:UART:BAUDrate command selects the bit rate (in bps) for
the serial decoder and/or trigger when in UART mode. The baud rate can
be set from 1200 b/s to 8 Mb/s in 100 b/s increments. If you enter a baud
rate that is not divisible by 100 b/s, the baud rate is set to the nearest
baud rate divisible by 100 b/s.
If the baud rate you select does not match the system baud rate, false
triggers may occur.
Query Syntax
:SBUS<n>:UART:BAUDrate?
The :SBUS<n>:UART:BAUDrate? query returns the current UART baud rate
setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 1200 to 8000000 in 100 b/s increments
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:MODE" on page 648
":SBUS<n>:UART:TRIGger:TYPE" on page 555