Agilent Technologies InfiniiVision 3000 DSO-X 3054A Servisní příručka Strana 396

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 970
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 395
396 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
20 :MTESt Commands
:MTESt:COUNt:FWAVeforms
(see page 860)
Query Syntax
:MTESt:COUNt:FWAVeforms? [CHANnel<n>]
<n> ::= 1 to (# analog channels) in NR1 format
The :MTESt:COUNt:FWAVeforms? query returns the total number of failed
waveforms in the current mask test run. This count is for all regions and
all waveforms.
Return Format
<failed><NL>
<failed> ::= number of failed waveforms in NR1 format.
See Also "Introduction to :MTESt Commands" on page 387
":MTESt:COUNt:WAVeforms" on page 399
":MTESt:COUNt:TIME" on page 398
":MTESt:COUNt:RESet" on page 397
":MTESt:SOURce" on page 417
Example Code "Example Code" on page 387
Zobrazit stránku 395
1 2 ... 391 392 393 394 395 396 397 398 399 400 401 ... 969 970

Komentáře k této Příručce

Žádné komentáře