Agilent Technologies InfiniiVision 3000 DSO-X 3054A Servisní příručka Strana 750

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 970
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 749
750 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
30 :WGEN Commands
:WGEN:FUNCtion:SQUare:DCYCle
(see page 860)
Command Syntax
:WGEN:FUNCtion:SQUare:DCYCle <percent>
<percent> ::= duty cycle percentage from 20% to 80% in NR3 format
For Square waveforms, the :WGEN:FUNCtion:SQUare:DCYCle command
specifies the square wave duty cycle.
Duty cycle is the percentage of the period that the waveform is high.
Query Syntax
:WGEN:FUNCtion:SQUare:DCYCle?
The :WGEN:FUNCtion:SQUare:DCYCle? query returns the currently set
square wave duty cycle.
Return Format
<percent><NL>
<percent> ::= duty cycle percentage from 20% to 80% in NR3 format
See Also "Introduction to :WGEN Commands" on page 744
":WGEN:FUNCtion" on page 746
Zobrazit stránku 749
1 2 ... 745 746 747 748 749 750 751 752 753 754 755 ... 969 970

Komentáře k této Příručce

Žádné komentáře