:WGEN Commands 30
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 757
:WGEN:VOLTage:LOW
(see page 860)
Command Syntax
:WGEN:VOLTage:LOW <low>
<low> ::= low-level voltage in volts, in NR3 format
For Sine, Square, Ramp, Pulse, and Noise waveforms, the
:WGEN:VOLTage:LOW command specifies the waveform's low- level voltage.
Use the :WGEN:VOLTage:HIGH command to specify the high- level voltage.
You can also specify the high- level and low- level voltages indirectly using
the :WGEN:VOLTage and :WGEN:VOLTage:OFFSet commands. For example,
a high- level voltage of 4 V and a low- level voltage of - 1 V is the same as
an amplitude of 5 V and an offset of 1 V.
Query Syntax
:WGEN:VOLTage:LOW?
The :WGEN:VOLTage:LOW? query returns the currently specified waveform
low- level voltage.
Return Format
<low><NL>
<low> ::= low-level voltage in volts, in NR3 format
See Also "Introduction to :WGEN Commands" on page 744
":WGEN:FUNCtion" on page 746
":WGEN:VOLTage:LOW" on page 757
":WGEN:VOLTage" on page 755
":WGEN:VOLTage:OFFSet" on page 758