Agilent Technologies InfiniiVision 3000 DSO-X 3054A Servisní příručka Strana 842

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 970
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 841
842 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
34 Status Reporting
Operation Status Event Register (:OPERegister[:EVENt])
The Operation Status Event Register register hosts these bits:
If any of these bits are set, the OPER bit (bit 7) of the Status Byte
Register is set. The Operation Status Event Register is read and cleared
with the :OPERegister[:EVENt]? query. The register output is enabled or
disabled using the mask value supplied with the OPEE command.
Name Location Description
RUN bit bit 3 Is set whenever the instrument goes from a stop
state to a single or running state.
WAIT TRIG bit bit 5 Is set by the Trigger Armed Event Register and
indicates that the trigger is armed.
MTE bit bit 9 Comes from the Mask Test Event Registers.
OVLR bit bit 11 Is set whenever a 50 input overload occurs.
Zobrazit stránku 841
1 2 ... 837 838 839 840 841 842 843 844 845 846 847 ... 969 970

Komentáře k této Příručce

Žádné komentáře