Agilent Technologies InfiniiVision 3000 DSO-X 3054A Servisní příručka Strana 646

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 970
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 645
646 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
28 :TRIGger Commands
:TRIGger:LEVel:HIGH
(see page 860)
Command Syntax
:TRIGger:LEVel:HIGH <level>, <source>
<level> ::= 0.75 x full-scale voltage from center screen in NR3 format
for internal triggers
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels) in NR1 format
The :TRIGger:LEVel:HIGH command sets the high trigger voltage level
voltage for the specified source.
High and low trigger levels are used with runt triggers and rise/fall time
(transition) triggers.
Query Syntax
:TRIGger:LEVel:HIGH? <source>
The :TRIGger:LEVel:HIGH? query returns the high trigger voltage level for
the specified source.
Return Format
<level><NL>
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:LEVel:LOW" on page 647
":TRIGger:RUNT Commands" on page 680
":TRIGger:TRANsition Commands" on page 691
":TRIGger[:EDGE]:SOURce" on page 661
Zobrazit stránku 645
1 2 ... 641 642 643 644 645 646 647 648 649 650 651 ... 969 970

Komentáře k této Příručce

Žádné komentáře