24 :SBUS<n> Commands
:SBUS<n>:DISPlay
(see page 860)
Command Syntax
:SBUS<n>:DISPlay <display>
<display> ::= {{1 | ON} | {0 | OFF}}
The :SBUS<n>:DISPlay command turns displaying of the serial decode bus
on or off.
Query Syntax
:SBUS<n>:DISPlay?
The :SBUS<n>:DISPlay? query returns the current display setting of the
serial decode bus.
Return Format
<display><NL>
<display> ::= {0 | 1}
Errors • "- 241, Hardware missing" on page 823
See Also • "Introduction to :SBUS<n> Commands" on page 453
• ":CHANnel<n>:DISPlay" on page 226
• ":DIGital<d>:DISPlay" on page 249
• ":POD<n>:DISPlay" on page 421
• ":VIEW" on page 186
• ":BLANk" on page 162
• ":STATus" on page 183
This command is only valid when a serial decode option has been licensed.
Two I2S buses or two SPI buses cannot be decoded on both SBUS1 and SBUS2 at the same
time.