AAgilent 1100 Series Diode Array and Multiple Wavelength DetectorsReference Manual

Strany 2

10 1100 Series DAD and MWD Reference ManualUsing the EMF Counters 235Electronics 237 Detector Main Board (DAM/MWM) 238Firmware Description 246Firmware

Strany 3 - In This Guide…

100 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorFigure 29 Main AssembliesFlow cell,see page 105 and page 108Leak sensor,see page 1

Strany 4

1100 Series DAD and MWD Reference Manual 101Repairing the Detector 4Simple RepairsOn the following pages repairs are described that can be carried out

Strany 5 - Contents

102 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorExchanging a Lamp When required If noise or drift exceeds application limits or la

Strany 6

1100 Series DAD and MWD Reference Manual 103Repairing the Detector 42Disconnect lamp from the connector and unscrew the lamp.3 Remove the lamp. Do not

Strany 7

104 1100 Series DAD and MWD Reference Manual4 Repairing the Detector6Replace the front cover. Next steps:• Reset the lamp counter as described in the

Strany 8

1100 Series DAD and MWD Reference Manual 105Repairing the Detector 4Exchanging a Flow Cell When required If an application needs a different type of f

Strany 9

106 1100 Series DAD and MWD Reference Manual4 Repairing the Detector2Press the release button and open the flow cell door. 3 Disconnect the cell inlet

Strany 10

1100 Series DAD and MWD Reference Manual 107Repairing the Detector 45Insert the flow cell into the cell compartment and close the cell door.6 Reconnec

Strany 11

108 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorRepairing the Standard or Semi-Micro Flow Cell When required If the flow cell need

Strany 12

1100 Series DAD and MWD Reference Manual 109Repairing the Detector 4Note:If you want to replace the gasket only, continue with step 8.Do not mix the g

Strany 13 - Installing the Detector

1100 Series DAD and MWD Reference Manual 11Screens available from the System screen 282Screens available from the Records screen 284Diagnostics and Te

Strany 14 - Site Requirements

110 1100 Series DAD and MWD Reference Manual4 Repairing the Detector4Assemble the washers and the window assembly in correct order.5 Correct orientati

Strany 15 - °C/hour (1.8 °F/hour)

1100 Series DAD and MWD Reference Manual 111Repairing the Detector 48Insert a new gasket [6, 7] and the window assembly [1] into the cell body. Assure

Strany 16 - Physical Specifications

112 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorRepairing the High Pressure Flow CellWhen required If the flow cell needs repair d

Strany 17 - Unpacking the Detector

1100 Series DAD and MWD Reference Manual 113Repairing the Detector 4If you want to replace the gasket only, continue with step 8.2 Use a tooth pick to

Strany 18 - Figure 1 Waste Tubing Parts

114 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorReplacing Capillaries on a Standard Flow CellWhen required If the capillary is blo

Strany 19 - Installing the Detector 1

1100 Series DAD and MWD Reference Manual 115Repairing the Detector 4NotesThe fittings at the flow cell body are special types for low dead volumes and

Strany 20

116 1100 Series DAD and MWD Reference Manual4 Repairing the Detector5 Unscrew the fixing screw and unwrap the inlet capillary from the grove in the fl

Strany 21

1100 Series DAD and MWD Reference Manual 117Repairing the Detector 49 The capillary lays in the grove and should be tied around the body (in the grove

Strany 22

118 1100 Series DAD and MWD Reference Manual4 Repairing the Detector13 Fix the new heat exchanger to the clamp unit.Next steps:• Reconnect the capilla

Strany 23 - Status indicator

1100 Series DAD and MWD Reference Manual 119Repairing the Detector 4Replacing Capillaries on a Semi-Micro and High Pressure Flow CellWhen required If

Strany 24

12 1100 Series DAD and MWD Reference Manual

Strany 25

120 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorNotesThe fittings at the flow cell body are special types for low dead volumes and

Strany 26

1100 Series DAD and MWD Reference Manual 121Repairing the Detector 45 Fix the new heat exchanger to the clamp unit and the heat exchanger to the cell

Strany 27

122 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorNano Flow Cell - Replacing or CleaningWhen required If parts are contaminated or l

Strany 28

1100 Series DAD and MWD Reference Manual 123Repairing the Detector 43 Unscrew the capillaries from the flow cell. DO NOT use the adapter at this time!

Strany 29 - How to optmize the Detector

124 1100 Series DAD and MWD Reference Manual4 Repairing the Detector7 Replace the cell seal assembly onto the quartz body. Always use a new seal assem

Strany 30

1100 Series DAD and MWD Reference Manual 125Repairing the Detector 411 Re-install the flow cell and connect the capillaries to the union holder.12 Per

Strany 31 - Optimization Overview

126 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorFigure 30 Wrench plus Torque AdapterWrenchAdapterDO NOT press down more than shown

Strany 32 - Figure 7 Choosing a Flow Cell

1100 Series DAD and MWD Reference Manual 127Repairing the Detector 4Cleaning or Exchanging the Holmium Oxide Filter When required If holmium oxide fi

Strany 33 - Flow Cell Path Length

128 1100 Series DAD and MWD Reference Manual4 Repairing the Detector2If not already in this position, move the filter up. 3 While releasing the holder

Strany 34

1100 Series DAD and MWD Reference Manual 129Repairing the Detector 45Replace the flow cell cover and fix the six screws. Next steps:• Perform a holmiu

Strany 35 - Peak width (response time)

13Agilent 1100 Series Diode Array and Multiple Wavelength DetectorsReference ManualAgilent Technologies1Installing the DetectorSite Requirements 14Phy

Strany 36

130 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorCorrecting Leaks1 Remove the front cover.2 Use tissue to dry the leak sensor area

Strany 37

1100 Series DAD and MWD Reference Manual 131Repairing the Detector 4Replacing Leak Handling System Parts1 Remove the front cover.2 Pull the leak funne

Strany 38 - Bandwidth

132 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorCleaning or Replacing Coupling Lens Assembly When required The coupling lens asse

Strany 39 - Slit Width

1100 Series DAD and MWD Reference Manual 133Repairing the Detector 42Carefully slide the lens assembly to the right into the deuterium lamp area by pu

Strany 40

134 1100 Series DAD and MWD Reference Manual4 Repairing the Detector6Insert the rebuilt or new lens assembly and slide it into its holder.7 Align the

Strany 41

1100 Series DAD and MWD Reference Manual 135Repairing the Detector 4Cleaning or Replacing Cell Support Window When required This window is located be

Strany 42 - Options for storing spectra

136 1100 Series DAD and MWD Reference Manual4 Repairing the Detector2If not already in this position, move the holmium oxide filter down.3 Use the 1.5

Strany 43 - Threshold

1100 Series DAD and MWD Reference Manual 137Repairing the Detector 46The assembly may be disassembled using a pair of pliers to remove the retainer ri

Strany 44 - Optimizing Selectivity

138 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorExchanging Internal PartsWARNINGThe following procedures require opening the main

Strany 45

1100 Series DAD and MWD Reference Manual 139Repairing the Detector 4WARNINGEye damage may result from directly viewing the light produced by the deute

Strany 46

14 1100 Series DAD and MWD Reference Manual1 Installing the DetectorSite RequirementsA suitable environment is important to ensure optimal performance

Strany 47

140 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorRemoving the Top Cover and Foam When required For all repairs inside the detector

Strany 48

1100 Series DAD and MWD Reference Manual 141Repairing the Detector 42Install the ESD strap. If installed, unscrew and remove the interface board. Plac

Strany 49

142 1100 Series DAD and MWD Reference Manual4 Repairing the Detector6Disconnect the lamps from their connectors. 7 Remove the two lamp connectors towa

Strany 50 - Test Funct io ns

1100 Series DAD and MWD Reference Manual 143Repairing the Detector 4Exchanging the Detector Main Board1 Turn off the lamp.2 Switch off the module, and

Strany 51

144 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorNOTEWhen removing connectors, counter-hold with one hand on connector J13.Figure 3

Strany 52

1100 Series DAD and MWD Reference Manual 145Repairing the Detector 47 Remove the detector main board. Place the board on the ESD kit. 8 In most cases

Strany 53

146 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorEntering the Serial Number using the Control Module1 Connect the control module to

Strany 54 - Timeout

1100 Series DAD and MWD Reference Manual 147Repairing the Detector 4Entering the Serial Number using the Agilent ChemStationModule serial numbers are

Strany 55 - Shutdown

148 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorExchanging the Fan (G1315A/G1365A)NOTEUse this procedure for the the G1315A/G1365A

Strany 56 - Remote Timeout

1100 Series DAD and MWD Reference Manual 149Repairing the Detector 4The fan must be installed in the correct orientation (see arrow on the fan) to ens

Strany 57 - Synchronization Lost

1100 Series DAD and MWD Reference Manual 15Installing the Detector 1Bench SpaceThe detector dimensions and weight (see Table 1) allows you to place th

Strany 58

150 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorReplacing Fan Parts (G1315B/G1365B)NOTEWith the introduction of the G1315B/G1365B

Strany 59

1100 Series DAD and MWD Reference Manual 151Repairing the Detector 4The heater and temperature sensor assemblies are clipped into the fan assembly. In

Strany 60 - Leak Sensor Open

152 1100 Series DAD and MWD Reference Manual4 Repairing the Detector5 To re-install the sensor holder (here shown with the sensor), bring the holder i

Strany 61 - Leak Sensor Short

1100 Series DAD and MWD Reference Manual 153Repairing the Detector 4The figure below shows the required air flow direction.8 Insert the fan assembly i

Strany 62 - Compensation Sensor Open

154 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorRemoving the Optical Unit (G1315A/G1365A)1 Turn off the lamp.2 Switch off detector

Strany 63 - Compensation Sensor Short

1100 Series DAD and MWD Reference Manual 155Repairing the Detector 410 Remove the two rubber shock absorbers. 11 Disconnect the SCI cable (J23) and th

Strany 64 - Fan Failed

156 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorRemoving the Optical Unit (G1315B/G1365B)When required For all repairs inside the

Strany 65 - Open Cover

1100 Series DAD and MWD Reference Manual 157Repairing the Detector 42 Remove the two rubber shock absorbers. 3 Slide the z-plane to the front and remo

Strany 66 - Cover Violation

158 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorExchanging the Source Lens (Achromat) AssemblyWhen required The source lens (achro

Strany 67 - Visible Lamp Current

1100 Series DAD and MWD Reference Manual 159Repairing the Detector 42 Pull the filter motor off the shaft. Keep the spring and the lever fixture in a

Strany 68 - Visible Lamp Voltage

16 1100 Series DAD and MWD Reference Manual1 Installing the DetectorPhysical SpecificationsTable 1 Physical SpecificationsType Specification CommentsW

Strany 69 - Communication Error

160 1100 Series DAD and MWD Reference Manual4 Repairing the Detector6 Remove the achromat assembly. 7 Use a metal blade (12-mm width, 1-mm thick) to u

Strany 70 - Calibration Values Invalid

1100 Series DAD and MWD Reference Manual 161Repairing the Detector 412 Before installing the filter motor, press the filter lever in from the other si

Strany 71 - Diode Current Leakage

162 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorReplacing the Holmium Oxide Filter MotorWhen required If defectiveTools required S

Strany 72 - Holmium Oxide Test Failed

1100 Series DAD and MWD Reference Manual 163Repairing the Detector 42 Pull the filter motor off the shaft. Keep the spring in a safe place.3 Before in

Strany 73 - UV Lamp Current

164 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorExchanging the Power Supply1 Turn off the lamp.2 Switch off the detector, and disc

Strany 74 - UV Lamp Voltage

1100 Series DAD and MWD Reference Manual 165Repairing the Detector 45 Carefully remove the bottom foam part by sliding it out towards the rear.6 Unscr

Strany 75 - UV Ignition Failed

166 1100 Series DAD and MWD Reference Manual4 Repairing the Detector9 Insert the power supply into its location and fix it with the screws at the rear

Strany 76 - UV Heater Current

1100 Series DAD and MWD Reference Manual 167Repairing the Detector 4Exchanging the Leak Sensor1 Turn off the lamp.2 Switch off the detector and discon

Strany 77

168 1100 Series DAD and MWD Reference Manual4 Repairing the Detector12 Remove the leak pan from the cabinet bottom. 13 Remove the leak sensor assembly

Strany 78 - Module Type Changed

1100 Series DAD and MWD Reference Manual 169Repairing the Detector 416 Replace the leak pan into the cabinet bottom.Next steps:• Reconnect the leak se

Strany 79 - Type B, Parameter 1 or 3

1100 Series DAD and MWD Reference Manual 17Installing the Detector 1Unpacking the DetectorDamaged PackagingIf the delivery packaging shows signs of ex

Strany 80 - Assembly

170 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorReplacing Status Light PipeWhen required If part was broken or removedTools requir

Strany 81

1100 Series DAD and MWD Reference Manual 171Repairing the Detector 4Installing the Optical Unit (G1315A/G1365A)When required When repairs have been co

Strany 82 - Heater Failed

172 1100 Series DAD and MWD Reference Manual4 Repairing the Detector2 Carefully insert the optical unit into the instrument holding the one hand at ba

Strany 83 - Heater Power At Limit

1100 Series DAD and MWD Reference Manual 173Repairing the Detector 4Installing the Optical Unit (G1315B/G1365B)When required When repairs have been co

Strany 84

174 1100 Series DAD and MWD Reference Manual4 Repairing the Detector2 Carefully insert the optical unit into the instrument holding the one hand at ba

Strany 85 - Test Functions

1100 Series DAD and MWD Reference Manual 175Repairing the Detector 4Replacing the Foam and the Top Cover When required When all repairs have been com

Strany 86 - Holmium Oxide Test

176 1100 Series DAD and MWD Reference Manual4 Repairing the Detector2Insert the foam and make sure that the foam is installed correctly.3 Press the la

Strany 87 - Test Failed

1100 Series DAD and MWD Reference Manual 177Repairing the Detector 46Replace the cover. 7 If required, insert the interface board and fix the screws.8

Strany 88 - Intensity Test

178 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorAssembling the Main CoverWhen required If cover is brokenTools required NoneParts

Strany 89 - Intensity Test Evaluation

1100 Series DAD and MWD Reference Manual 179Repairing the Detector 4Replacing the Interface BoardWhen required For all repairs inside the detector or

Strany 90 - Cell Test

18 1100 Series DAD and MWD Reference Manual1 Installing the DetectorDetector Accessory Kit ContentsTable 3 Accessory Kit Contents (Part Number G1315-6

Strany 91 - Dark-Current Test

180 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorReplacing the Detector’s FirmwareThe installation of new firmware is required:• if

Strany 92

181Agilent 1100 Series Diode Array and Multiple Wavelength DetectorsReference ManualAgilent Technologies5Identifying Parts and MaterialsOverview of Ma

Strany 93 - Filter Test

182 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsOverview of Main AssembliesFigure 35 Overview of Main Assemblies6417238 1

Strany 94

1100 Series DAD and MWD Reference Manual 183Identifying Parts and Materials 5Table 8 Main AssembliesItem Description Part Number1 Main board DAM for G

Strany 95 - Repairing the Detector

184 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsOptical Unit AssemblyTable 9 Optical Unit AssemblyItem Description Part N

Strany 96

1100 Series DAD and MWD Reference Manual 185Identifying Parts and Materials 5Figure 36 Optical Unit Parts15436721481516711910121317

Strany 97 - Cleaning the Detector

186 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsStandard Flow CellTable 1 0 Standard Flow Cell PartsItem Description Part

Strany 98 - Using the ESD Strap

1100 Series DAD and MWD Reference Manual 187Identifying Parts and Materials 5Figure 37 Standard Flow Cell PartsFigure 38 Orientation of Spring Washers

Strany 99

188 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsSemi-Micro Flow CellTable 1 1 Semi-Micro Flow Cell PartsItem Description

Strany 100 - Figure 29 Main Assemblies

1100 Series DAD and MWD Reference Manual 189Identifying Parts and Materials 5Figure 39 Semi-Micro Flow Cell PartsFigure 40 Orientation of Spring Washe

Strany 101 - Simple Repairs

1100 Series DAD and MWD Reference Manual 19Installing the Detector 1Figure 2 Inlet Capillary (Column-Detector) PartsFerrule frontThis side is preinsta

Strany 102 - Exchanging a Lamp

190 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsHigh Pressure Flow CellTable 1 2 Repair Parts: Micro Flow Cell AssemblyIt

Strany 103

1100 Series DAD and MWD Reference Manual 191Identifying Parts and Materials 5Figure 41 High Pressure Flow Cell Assembly Parts1232456789

Strany 104 - • Turn the lamp on

192 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsNano Flow CellsThe following kits are available:Figure 42 shows all parts

Strany 105 - Exchanging a Flow Cell

1100 Series DAD and MWD Reference Manual 193Identifying Parts and Materials 5Table 14 lists the generic parts for both nano-flow cells:Table 1 4 Gener

Strany 106

194 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsTable 15 lists the specific parts for the 500 nl flow cell.Table 1 5 Spec

Strany 107 - • Replace the front cover

1100 Series DAD and MWD Reference Manual 195Identifying Parts and Materials 5Table 16 lists the specific parts for the 80 nl flow cell.Table 1 6 Speci

Strany 108 - • Remove the front cover

196 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsFan Assembly PartsTable 1 7 Fan Assembly PartsItem Description Part Numbe

Strany 109

1100 Series DAD and MWD Reference Manual 197Identifying Parts and Materials 5Holmium Oxide FilterTable 1 8 Holmium Oxide Filter Assembly PartsItem Des

Strany 110

198 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsCoupling Lens AssemblyTable 1 9 Coupling Lens AssemblyItem Description Pa

Strany 111 - • Insert the flow cell

1100 Series DAD and MWD Reference Manual 199Identifying Parts and Materials 5Source Lens (Achromat) AssemblyTable 2 0 Source Lens AssemblyItem Descrip

Strany 112

2 1100 Series DAD and MWD Reference ManualNotices© Agilent Technologies, Inc. 1995, 1996-2004No part of this manual may be reproduced in any form or b

Strany 113

20 1100 Series DAD and MWD Reference Manual1 Installing the DetectorOptimizing the Stack ConfigurationIf your detector is part of a complete Agilent 1

Strany 114

200 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsCell Support AssemblyTable 2 1 Cell Support AssemblyItem Description Part

Strany 115

1100 Series DAD and MWD Reference Manual 201Identifying Parts and Materials 5Control ModuleTable 2 2 Cotrol Module PartsItem Description Part NumberCo

Strany 116

202 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsSheet Metal Kit (G1315A/G1365A)NOTEFor the information on the sheet metal

Strany 117 - Repairing the Detector 4

1100 Series DAD and MWD Reference Manual 203Identifying Parts and Materials 5Sheet Metal Kit (G1315B/G1365B)NOTEFor the information on the sheet metal

Strany 118 - Next steps:

204 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsPlastic PartsTable 2 5 Plastics PartsItem Description Part Number1 Front

Strany 119

1100 Series DAD and MWD Reference Manual 205Identifying Parts and Materials 5Foam PartsTable 2 6 Foam PartsItem Description Part Number1, 2 EPP foam k

Strany 120

206 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsPower and Status Light PipesTable 2 7 Power and Status Light PipesItem De

Strany 121

1100 Series DAD and MWD Reference Manual 207Identifying Parts and Materials 5Leak PartsTable 2 8 Leak PartsItem Description Part Number1 Leak sensor a

Strany 122

208 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsAccessory KitThis kit contains some accessories and tools needed for the

Strany 123

1100 Series DAD and MWD Reference Manual 209Identifying Parts and Materials 5Figure 55 Waste Tubing PartsFigure 56 Inlet Capillary (Column-Detector) P

Strany 124

1100 Series DAD and MWD Reference Manual 21Installing the Detector 1Figure 4 Recommended Stack Configuration (Rear View)GPIB or LANto LC ChemStationAn

Strany 125

210 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsCable OverviewWARNINGNever use cables other than the ones supplied by Ali

Strany 126

1100 Series DAD and MWD Reference Manual 211Identifying Parts and Materials 5BCD cables3396 integrator 03396-60560General purpose (spade Lugs) G1351-8

Strany 127

212 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsAnalog CablesOne end of these cables provides a BNC connector to be conne

Strany 128

1100 Series DAD and MWD Reference Manual 213Identifying Parts and Materials 5Agilent 1100 to 3394/6 IntegratorsAgilent 1100 to BNC ConnectorConnector3

Strany 129 - • Turn on the flow

214 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsAgilent 1100 to General PurposeConnector01046-60105Pin3394/6PinAgilent 11

Strany 130 - Correcting Leaks

1100 Series DAD and MWD Reference Manual 215Identifying Parts and Materials 5Remote CablesOne end of these cables provides a Aligent Technologies APG

Strany 131

216 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsAgilent 1100 to 3392/3 IntegratorsAgilent 1100 to 3394 IntegratorsConnect

Strany 132 - • Turn off the lamp(s)

1100 Series DAD and MWD Reference Manual 217Identifying Parts and Materials 5Agilent 1100 to 3396A IntegratorsAgilent 1100 to 3396 Series II / 3395A I

Strany 133

218 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsAgilent 1100 to 3396 Series III / 3395B IntegratorsAgilent 1100 to HP 105

Strany 134 - • Replace both lamps

1100 Series DAD and MWD Reference Manual 219Identifying Parts and Materials 5Agilent 1100 to HP 1090 LC or Signal Distribution ModuleAgilent 1100 to G

Strany 135

22 1100 Series DAD and MWD Reference Manual1 Installing the DetectorInstalling the Detector1 Install the LAN interface board in the detector (if requi

Strany 136

220 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsBCD CablesOne end of these cables provides a 15-pin BCD connector to be c

Strany 137

1100 Series DAD and MWD Reference Manual 221Identifying Parts and Materials 5Agilent 1100 to 3396 IntegratorsConnector03396-60560Pin3392/3PinAgilent 1

Strany 138 - Exchanging Internal Parts

222 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsAuxiliary CableOne end of this cable provides a modular plug to be connec

Strany 139

1100 Series DAD and MWD Reference Manual 223Identifying Parts and Materials 5CAN CableBoth ends of this cable provide a modular plug to be connected t

Strany 140 - • Disconnect capillaries

224 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsExternal Contact CableOne end of this cable provides a 15-pin plug to be

Strany 141

1100 Series DAD and MWD Reference Manual 225Identifying Parts and Materials 5RS-232 Cable KitThis kit contains a 9-pin female to 9-pin female Null Mod

Strany 142

226 1100 Series DAD and MWD Reference Manual5 Identifying Parts and MaterialsLAN CablesRecommended CablesTable 3 1Description Part numberCross-over ne

Strany 143

227Agilent 1100 Series Diode Array and Multiple Wavelength DetectorsReference ManualAgilent Technologies6Introduction to the DetectorIntroduction to t

Strany 144

228 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorIntroduction to the DetectorThe detector is designed for highest optical per

Strany 145

1100 Series DAD and MWD Reference Manual 229Introduction to the Detector 6Optical System OverviewOptical SystemThe optical system of the detector is s

Strany 146

1100 Series DAD and MWD Reference Manual 23Installing the Detector 14 Connect the power cable to the power connector at the rear of the detector.5 Con

Strany 147

230 1100 Series DAD and MWD Reference Manual6 Introduction to the Detectorwavelength range. The light source for the visible and SWNIR wavelength rang

Strany 148

1100 Series DAD and MWD Reference Manual 231Introduction to the Detector 6Diode Array The diode array is a series of 1024 individual photodiodes and c

Strany 149 - AIRFLOW

232 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorElectrical Connections• The GPIB connector is used to connect the detector w

Strany 150

1100 Series DAD and MWD Reference Manual 233Introduction to the Detector 6Figure 58 Electrical ConnectionsAnalog signalsAPG remoteRS-232CCANGPIB Power

Strany 151

234 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorInstrument LayoutThe industrial design of the detector incorporates several

Strany 152

1100 Series DAD and MWD Reference Manual 235Introduction to the Detector 6Early Maintenance Feedback (EMF)Maintenance requires the exchange of compone

Strany 153 - G1315/65B

236 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorSetting the EMF LimitsThe setting of the EMF limits must be optimized over o

Strany 154

1100 Series DAD and MWD Reference Manual 237Introduction to the Detector 6ElectronicsThe electronics are comprised of four main components:• detector

Strany 155

238 1100 Series DAD and MWD Reference Manual6 Introduction to the Detector Detector Main Board (DAM/MWM)This board controls all information and activi

Strany 156

1100 Series DAD and MWD Reference Manual 239Introduction to the Detector 6Conversion and subsequent calculation to achieve absorbance values include t

Strany 157

24 1100 Series DAD and MWD Reference Manual1 Installing the DetectorFigure 6 Rear View of DetectorAnalog signalAPG remoteRS-232CCANGPIBPowerConfigurat

Strany 158

240 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorFigure 59 Block Diagram Detector Main Controller FunctionalityMain processor

Strany 159

1100 Series DAD and MWD Reference Manual 241Introduction to the Detector 6SCI BoardThe SCI (spectrograph connector interface), located on the optical

Strany 160

242 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorFigure 60 Block Diagram PFP FunctionsPFP ASIC EPLDRAMTest signals, GND, Vref

Strany 161

1100 Series DAD and MWD Reference Manual 243Introduction to the Detector 6Deuterium Lamp Filament ControlBefore ignition, the deuterium lamp filament

Strany 162

244 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorFigure 61 Block Diagram Lamp Supply FunctionsDeuterium lamp filament control

Strany 163

1100 Series DAD and MWD Reference Manual 245Introduction to the Detector 6Diagnostic A/D ConverterThe diagnostic A/D converter senses currents and vol

Strany 164 - Exchanging the Power Supply

246 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorFirmware DescriptionThe firmware of the instrument consists of two independe

Strany 165

1100 Series DAD and MWD Reference Manual 247Introduction to the Detector 6Firmware UpdatesFirmware updates can be done using your user interface:• han

Strany 166

248 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorRaw Data Conversion to AbsorbanceThe raw data flow (from the photodiode arra

Strany 167 - Exchanging the Leak Sensor

1100 Series DAD and MWD Reference Manual 249Introduction to the Detector 6Figure 63 Firmware Flow DiagramTem pe rat ure filterRaw dataTemperatureSubtr

Strany 168

1100 Series DAD and MWD Reference Manual 25Installing the Detector 1Flow Connections to the DetectorPreparations Detector is installed in the LC syste

Strany 169

250 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorTemperature CorrectionThe quantum efficiency of the photodiode array depends

Strany 170 - Replacing Status Light Pipe

1100 Series DAD and MWD Reference Manual 251Introduction to the Detector 6continuous, linear scale, an interpolation algorithm is applied. This algori

Strany 171

252 1100 Series DAD and MWD Reference Manual6 Introduction to the Detectorslope of this signal (upslope, apex, downslope and baseline) and stores spec

Strany 172

1100 Series DAD and MWD Reference Manual 253Introduction to the Detector 6Optional Interface BoardsThe Agilent 1100 Series modules have one optional b

Strany 173

254 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorThere are general purpose cables available to connect the BCD output, see “B

Strany 174

1100 Series DAD and MWD Reference Manual 255Introduction to the Detector 6LAN Communication Interface BoardThe following cards can be used with the Ag

Strany 175

256 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorInterfacesThe Agilent 1100 Series modules provide the following interfaces:•

Strany 176

1100 Series DAD and MWD Reference Manual 257Introduction to the Detector 6Analog Signal OutputThe analog signal output can be distributed to a recordi

Strany 177

258 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorRemote control allows easy connection between single instruments or systems

Strany 178 - Assembling the Main Cover

1100 Series DAD and MWD Reference Manual 259Introduction to the Detector 6RS-232CThe RS-232C connector is used to control the instrument from a comput

Strany 179 - Replacing the Interface Board

26 1100 Series DAD and MWD Reference Manual1 Installing the Detector1 Press the release buttons and remove the front cover to gain access to the flow

Strany 180

260 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorFigure 65 RS-232 CableInstrumentDCDRXTXDTRGNDDSRRTSCTSRIDCDRXTXDTRGNDDSRRTSC

Strany 181 - Agilent Technologies

1100 Series DAD and MWD Reference Manual 261Introduction to the Detector 6Setting the 8-bit Configuration SwitchThe 8-bit configuration switch is loca

Strany 182 - Overview of Main Assemblies

262 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorIn the non-volatile memory, the parameters are kept, regardless of whether y

Strany 183 - Table 8 Main Assemblies

1100 Series DAD and MWD Reference Manual 263Introduction to the Detector 6Communication Settings for RS-232C CommunicationThe communication protocol u

Strany 184 - Optical Unit Assembly

264 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorOne start bit and one stop bit are always used (not selectable).Per default,

Strany 185 - Figure 36 Optical Unit Parts

1100 Series DAD and MWD Reference Manual 265Introduction to the Detector 6Stay-Resident SettingsFirmware update procedures may require this mode in ca

Strany 186 - Standard Flow Cell

266 1100 Series DAD and MWD Reference Manual6 Introduction to the DetectorThe Main Power Supply AssemblyThe main power supply comprises a closed assem

Strany 187

1100 Series DAD and MWD Reference Manual 267Introduction to the Detector 6No accessible hardware fuse is needed because the main power supply is safe

Strany 188 - Semi-Micro Flow Cell

268 1100 Series DAD and MWD Reference Manual6 Introduction to the Detector

Strany 189

269Agilent 1100 Series Diode Array and Multiple Wavelength DetectorsReference ManualAgilent Technologies7Control Module Screens for Agilent 1100 DAD a

Strany 190 - High Pressure Flow Cell

1100 Series DAD and MWD Reference Manual 27Installing the Detector 15 Assemble the column-detector capillary from the accessory kit. One side is alrea

Strany 191

270 1100 Series DAD and MWD Reference Manual7 Control Module Screens for Agilent 1100 DAD and MWDMajor keys on the Agilent 1100 Control ModuleTa b l e

Strany 192 - Nano Flow Cells

1100 Series DAD and MWD Reference Manual 271Control Module Screens for Agilent 1100 DAD and MWD 7Control Module - Unsupported ModuleIn case the contro

Strany 193 - Table 1 4 Generic Parts

272 1100 Series DAD and MWD Reference Manual7 Control Module Screens for Agilent 1100 DAD and MWDScreens available from the Analysis screenThe Analysi

Strany 194

1100 Series DAD and MWD Reference Manual 273Control Module Screens for Agilent 1100 DAD and MWD 7The lamp(s) can be selected individually.Setup ViewIn

Strany 195

274 1100 Series DAD and MWD Reference Manual7 Control Module Screens for Agilent 1100 DAD and MWDHere, e.g. the column compartment parameters are show

Strany 196 - Fan Assembly Parts

1100 Series DAD and MWD Reference Manual 275Control Module Screens for Agilent 1100 DAD and MWD 7Use the m-key for the context sensitive menu. The Sta

Strany 197 - Holmium Oxide Filter

276 1100 Series DAD and MWD Reference Manual7 Control Module Screens for Agilent 1100 DAD and MWDSelecting More and Signals C-E on the pull down menu

Strany 198 - Coupling Lens Assembly

1100 Series DAD and MWD Reference Manual 277Control Module Screens for Agilent 1100 DAD and MWD 7Press the F7 key (Insert) to add a timetable events.

Strany 199

278 1100 Series DAD and MWD Reference Manual7 Control Module Screens for Agilent 1100 DAD and MWDSettings - Spectrum RangeFOR G1315A/B DAD: With the F

Strany 200 - Cell Support Assembly

1100 Series DAD and MWD Reference Manual 279Control Module Screens for Agilent 1100 DAD and MWD 7Press F5 key (Views) and select Status.StatusThis is

Strany 201 - Control Module

28 1100 Series DAD and MWD Reference Manual1 Installing the Detector9 Remove the flow cell and establish a flow and observe for leaks.10 Insert the fl

Strany 202

280 1100 Series DAD and MWD Reference Manual7 Control Module Screens for Agilent 1100 DAD and MWDSignal plotPress F6 key (Plot) to enter the plot scre

Strany 203

1100 Series DAD and MWD Reference Manual 281Control Module Screens for Agilent 1100 DAD and MWD 7Method screensUse ESC to return to the Analysis scree

Strany 204 - Plastic Parts

282 1100 Series DAD and MWD Reference Manual7 Control Module Screens for Agilent 1100 DAD and MWDScreens available from the System screenSystem screen

Strany 205

1100 Series DAD and MWD Reference Manual 283Control Module Screens for Agilent 1100 DAD and MWD 7System - ConfigurationUse the Esc key to receive View

Strany 206 - Power and Status Light Pipes

284 1100 Series DAD and MWD Reference Manual7 Control Module Screens for Agilent 1100 DAD and MWDScreens available from the Records screenRecords scre

Strany 207 - Leak Parts

1100 Series DAD and MWD Reference Manual 285Control Module Screens for Agilent 1100 DAD and MWD 7Info LogUse the m-key to receive a pop-up menu, Selec

Strany 208 - Accessory Kit

286 1100 Series DAD and MWD Reference Manual7 Control Module Screens for Agilent 1100 DAD and MWDFirmware UpdateUse the Esc key to receive Views on th

Strany 209 - Figure 55 Waste Tubing Parts

1100 Series DAD and MWD Reference Manual 287Control Module Screens for Agilent 1100 DAD and MWD 7Use the F5 key (FW Update) to enter the Update sectio

Strany 210 - Cable Overview

288 1100 Series DAD and MWD Reference Manual7 Control Module Screens for Agilent 1100 DAD and MWDChanging the serial numberIn case the serial number o

Strany 211

1100 Series DAD and MWD Reference Manual 289Control Module Screens for Agilent 1100 DAD and MWD 7Maintenance activitiesUse the Esc key to receive View

Strany 212 - Analog Cables

29Agilent 1100 Series Diode Array and Multiple Wavelength DetectorsReference ManualAgilent Technologies2How to optmize the DetectorOptimizing the Dete

Strany 213 - Agilent 1100 to BNC Connector

290 1100 Series DAD and MWD Reference Manual7 Control Module Screens for Agilent 1100 DAD and MWDDiagnostics and TestsTests screenUse the Esc key to r

Strany 214

1100 Series DAD and MWD Reference Manual 291Control Module Screens for Agilent 1100 DAD and MWD 7Wavelength CalibrationUse the F1 key (Calibrate) and

Strany 215 - Remote Cables

292 1100 Series DAD and MWD Reference Manual7 Control Module Screens for Agilent 1100 DAD and MWDBuilt-in Test Chromatogram (Simulation)To start the b

Strany 216

1100 Series DAD and MWD Reference Manual 293Control Module Screens for Agilent 1100 DAD and MWD 7Use the F6 key (Plot) to view the signal. A simulated

Strany 217

294 1100 Series DAD and MWD Reference Manual7 Control Module Screens for Agilent 1100 DAD and MWD

Strany 218

295Agilent 1100 Series Diode Array and Multiple Wavelength DetectorsReference ManualAgilent Technologies8SpecificationsPerformance Specifications 296

Strany 219

296 1100 Series DAD and MWD Reference Manual8 SpecificationsPerformance SpecificationsTable 5 0 Performance Specifications Agilent 1100 Series DAD and

Strany 220 - BCD Cables

1100 Series DAD and MWD Reference Manual 297Specifications 8Control and data evaluationAgilent ChemStation for LCAnalog outputs Recorder/integrator: 1

Strany 221

298 1100 Series DAD and MWD Reference Manual8 Specifications

Strany 222 - Auxiliary Cable

299Agilent 1100 Series Diode Array and Multiple Wavelength DetectorsReference ManualAgilent TechnologiesASafety InformationGeneral Safety Information

Strany 223 - CAN Cable

1100 Series DAD and MWD Reference Manual 3In This Guide…This manual contains information for using your Diode Array and Multiple Wavelength Detectors.

Strany 224 - External Contact Cable

30 1100 Series DAD and MWD Reference Manual2 How to optmize the DetectorOptimizing the Detector PerformanceThe detector has a variety of parameters th

Strany 225 - RS-232 Cable Kit

300 1100 Series DAD and MWD Reference ManualA Safety InformationGeneral Safety InformationThe following general safety precautions must be observed du

Strany 226 - LAN Cables

1100 Series DAD and MWD Reference Manual 301Safety Information ASome adjustments described in the manual, are made with power supplied to the instrume

Strany 227 - Introduction to the Detector

302 1100 Series DAD and MWD Reference ManualA Safety InformationSafety SymbolsTable 51 shows safety symbols used on the instrument and in the manuals.

Strany 228

1100 Series DAD and MWD Reference Manual 303Safety Information ALithium Batteries InformationWARNINGDanger of explosion if battery is incorrectly repl

Strany 229 - Optical System Overview

304 1100 Series DAD and MWD Reference ManualA Safety InformationDisposal of Mercury from Deuterium LampThe long-life deuterium lamp (part number 5181-

Strany 230

1100 Series DAD and MWD Reference Manual 305Safety Information ARadio InterferenceNever use cables other than the ones supplied by Aligent Technologie

Strany 231

306 1100 Series DAD and MWD Reference ManualA Safety InformationSound EmissionManufacturer’s DeclarationThis statement is provided to comply with the

Strany 232 - Electrical Connections

1100 Series DAD and MWD Reference Manual 307Safety Information AUV-RadiationEmissions of ultraviolet radiation (200-315 nm) from this product is limit

Strany 233

308 1100 Series DAD and MWD Reference ManualA Safety InformationSolvent InformationObserve the following recommendations on the use of solvents.Flow C

Strany 234 - Instrument Layout

1100 Series DAD and MWD Reference Manual 309Safety Information A• Chromatographic grade ethers, which can contain peroxides (for example, THF, dioxane

Strany 235 - Using the EMF Counters

1100 Series DAD and MWD Reference Manual 31How to optmize the Detector 2Optimization OverviewTable 4 Optimization OverviewParameter Impact1 Selection

Strany 236 - Setting the EMF Limits

310 1100 Series DAD and MWD Reference ManualA Safety InformationDeclaration of Conformity for HOX2 FilterWe herewith inform you that the holmium oxide

Strany 237 - Electronics

1100 Series DAD and MWD Reference Manual 311Safety Information AAgilent Technologies on InternetFor the latest information on products and services vi

Strany 238 - Main processor

312 1100 Series DAD and MWD Reference ManualA Safety Information

Strany 239 - On-board Battery

1100 Series DAD and MWD Reference Manual 313IndexAaccessory kit, 18, 208accuracy of wavelength, 296achromat (source lens), 229Agilent on internet, 311

Strany 240

314 1100 Series DAD and MWD Reference ManualIndexvisible lamp current, 67visible lamp voltage, 68wavelength calibration failed, 70ESD (electrostatic d

Strany 241 - Analog Outputs

1100 Series DAD and MWD Reference Manual 315Indexfor sensitivity, selectivity, linearity, dispersion, 33how to get the best performance, 30margins for

Strany 242

316 1100 Series DAD and MWD Reference ManualIndexspectrograph, 229, 230diodes per nm, 230stack configuration, 20, 21front view, 20rear view, 21steps,

Strany 244

 Agilent Technologies 1995, 1996-2004Printed in Germany Edition 05/2004*G1315-90005**G1315-90005*G1315-90005www.agilent.comAgilent TechnologiesIn Thi

Strany 245 - Temperature Control

32 1100 Series DAD and MWD Reference Manual2 How to optmize the Detector5 Setting the slit width• Use 4 nm slit for normal applications.• Use narrow s

Strany 246 - Firmware Description

1100 Series DAD and MWD Reference Manual 33How to optmize the Detector 2Optimizing for Sensitivity, Selectivity, Linearity and DispersionFlow Cell Pat

Strany 247 - Firmware Updates

34 1100 Series DAD and MWD Reference Manual2 How to optmize the DetectorTraditionally LC analysis with UV detectors is based on comparing measurements

Strany 248 - Gain Correction

1100 Series DAD and MWD Reference Manual 35How to optmize the Detector 2Peak width (response time)Response time describes how fast the detector signal

Strany 249

36 1100 Series DAD and MWD Reference Manual2 How to optmize the DetectorSample and Reference Wavelength and BandwidthThe detector measures absorbance

Strany 250

1100 Series DAD and MWD Reference Manual 37How to optmize the Detector 2Signal A in the detector default method is set to sample 250,100, reference 36

Strany 251 - Peak Detector

38 1100 Series DAD and MWD Reference Manual2 How to optmize the DetectorA wide bandwidth has the advantage of reducing noise by averaging over a wavel

Strany 252

1100 Series DAD and MWD Reference Manual 39How to optmize the Detector 2Slit WidthThe detector has a variable slit at the entrance of the spectrograph

Strany 253 - Optional Interface Boards

4 1100 Series DAD and MWD Reference Manual

Strany 254

40 1100 Series DAD and MWD Reference Manual2 How to optmize the DetectorA wide slit uses more of the light shining through the flow cell. This gives l

Strany 255 - Recommended Cables

1100 Series DAD and MWD Reference Manual 41How to optmize the Detector 2Furthermore, the absorbance is no longer strictly linear with concentration fo

Strany 256

42 1100 Series DAD and MWD Reference Manual2 How to optmize the DetectorOptimizing Spectral Acquisition (DAD only)Storage of all spectra consumes a lo

Strany 257 - Remote Interface

1100 Series DAD and MWD Reference Manual 43How to optmize the Detector 2ThresholdSets the peak detector. Only spectra from peaks higher than threshold

Strany 258

44 1100 Series DAD and MWD Reference Manual2 How to optmize the DetectorOptimizing SelectivityQuantifying Coeluting Peaks by Peak SuppressionIn chroma

Strany 259

1100 Series DAD and MWD Reference Manual 45How to optmize the Detector 2With a UV-visible detector based on a diode array and the correct choice of a

Strany 260 - Figure 65 RS-232 Cable

46 1100 Series DAD and MWD Reference Manual2 How to optmize the DetectorRatio Qualifiers for Selective Detection of Compound ClassesRatio qualifiers c

Strany 261

1100 Series DAD and MWD Reference Manual 47How to optmize the Detector 2In a four-component mixture, only biphenyl was recorded. The other three peaks

Strany 262 - GPIB Default Addresses

48 1100 Series DAD and MWD Reference Manual2 How to optmize the DetectorThe ratio range was set at 2 – 2.4 (2.2 ±10%). Only when the ratio between 249

Strany 263

49Agilent 1100 Series Diode Array and Multiple Wavelength DetectorsReference ManualAgilent Technologies3Troubleshooting and Test FunctionsOverview of

Strany 264 - Forced Cold Start Settings

1100 Series DAD and MWD Reference Manual 5Contents1 Installing the DetectorSite Requirements 14Physical Specifications 16Unpacking the Detector 17Dama

Strany 265 - Stay-Resident Settings

50 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsOverview of the Detector’s Indicators and Test FunctionsStatus Indicato

Strany 266

1100 Series DAD and MWD Reference Manual 51Troubleshooting and Test Functions 3Status IndicatorsTwo status indicators are located on the front of the

Strany 267

52 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test Functions• When the status indicator is OFF (and power switch light is on), the

Strany 268

1100 Series DAD and MWD Reference Manual 53Troubleshooting and Test Functions 3Error MessagesError messages are displayed in the user interface when a

Strany 269 - 1100 DAD and MWD

54 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsTimeout The timeout threshold was exceeded.Probable Causes• The analysi

Strany 270

1100 Series DAD and MWD Reference Manual 55Troubleshooting and Test Functions 3Shutdown An external instrument has generated a shut-down signal (throu

Strany 271

56 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsRemote Timeout A not-ready condition is still present on the remote inp

Strany 272 - Lamp ON/OFF

1100 Series DAD and MWD Reference Manual 57Troubleshooting and Test Functions 3Synchronization Lost During an analysis, the internal synchronization o

Strany 273 - Setup View

58 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsLeak A leak was detected in the detector.The signals from the two tempe

Strany 274 - Settings

1100 Series DAD and MWD Reference Manual 59Troubleshooting and Test Functions 3to port# 4 of the sampler switching valve. This tube can then be taken

Strany 275

6 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsOverview of the Detector’s Indicators and Test Functions 50Status Indica

Strany 276 - Settings - Timetable

60 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsLeak Sensor Open The leak sensor in the detector has failed (open circu

Strany 277 - Settings - Analog

1100 Series DAD and MWD Reference Manual 61Troubleshooting and Test Functions 3Leak Sensor Short The leak sensor in the detector has failed (short cir

Strany 278 - Settings - Run times

62 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsCompensation Sensor Open The ambient-compensation sensor (NTC) on the d

Strany 279

1100 Series DAD and MWD Reference Manual 63Troubleshooting and Test Functions 3Compensation Sensor Short The ambient-compensation sensor (NTC) on the

Strany 280 - Signal plot

64 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsFan Failed The cooling fan in the detector has failed. The hall sensor

Strany 281 - Method screens

1100 Series DAD and MWD Reference Manual 65Troubleshooting and Test Functions 3Open Cover The top foam has been removed.The sensor on the detector mai

Strany 282 - System - Control

66 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsCover Violation The top foam has been removed.The sensor on the detecto

Strany 283 - System - Configuration

1100 Series DAD and MWD Reference Manual 67Troubleshooting and Test Functions 3Visible Lamp Current The visible lamp current is missing. The processor

Strany 284 - System / Error Log

68 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsVisible Lamp Voltage The visible lamp voltage is missing. The processor

Strany 285 - Info Log

1100 Series DAD and MWD Reference Manual 69Troubleshooting and Test Functions 3Communication Error An internal communication error between the main pr

Strany 286 - Firmware Update

1100 Series DAD and MWD Reference Manual 7Illegal Value From „Air Inlet" Temperature Sensor 81Heater Failed 82Heater Power At Limit 83Wavelength

Strany 287

70 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsCalibration Values Invalid The calibration values read from the spectro

Strany 288 - Changing the serial number

1100 Series DAD and MWD Reference Manual 71Troubleshooting and Test Functions 3Diode Current Leakage When the detector is switched on, the processor c

Strany 289 - Maintenance activities

72 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsHolmium Oxide Test FailedThe holmium oxide test in the detector has fai

Strany 290 - Diagnostics and Tests

1100 Series DAD and MWD Reference Manual 73Troubleshooting and Test Functions 3UV Lamp Current The UV lamp current is missing. The processor continual

Strany 291 - Wavelength Calibration

74 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsUV Lamp VoltageThe UV lamp anode voltage is missing.The processor conti

Strany 292

1100 Series DAD and MWD Reference Manual 75Troubleshooting and Test Functions 3UV Ignition Failed The UV lamp failed to ignite. The processor monitors

Strany 293 - Slit Test

76 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsUV Heater CurrentThe UV lamp heater current is missing. During UV lamp

Strany 294

1100 Series DAD and MWD Reference Manual 77Troubleshooting and Test Functions 3New Messages with the G1315B and G1365B detectorsWith the introduction

Strany 295 - Specifications

78 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsModule Type ChangedAt switch-on, the detector has recognized a new conf

Strany 296 - Performance Specifications

1100 Series DAD and MWD Reference Manual 79Troubleshooting and Test Functions 3Type B, Parameter 1 or 3Probable Causes• Detector has recognized the ha

Strany 297

8 1100 Series DAD and MWD Reference ManualExchanging the Detector Main Board 143Entering the Serial Number using the Control Module 146Entering the Se

Strany 298 - 8 Specifications

80 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsIllegal Value From Temperature Sensor Mounted On The Fan AssemblyThis t

Strany 299 - Safety Information

1100 Series DAD and MWD Reference Manual 81Troubleshooting and Test Functions 3Illegal Value From „Air Inlet" Temperature SensorThis temperature

Strany 300 - General Safety Information

82 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsHeater FailedEvery time the deuterium lamp or the tungsten lamp is swit

Strany 301

1100 Series DAD and MWD Reference Manual 83Troubleshooting and Test Functions 3Heater Power At LimitThe available power of the heater reached either t

Strany 302 - Safety Symbols

84 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsWavelength RecalibrationThe detector uses the alpha (656.1 nm) and beta

Strany 303 - Lithium Batteries Information

1100 Series DAD and MWD Reference Manual 85Troubleshooting and Test Functions 3Test FunctionsHolmium Oxide TestThe built-in holmium oxide filter is us

Strany 304 - A Safety Information

86 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsHolmium Oxide TestThe holmium oxide test uses three characteristic abso

Strany 305 - Radio Interference

1100 Series DAD and MWD Reference Manual 87Troubleshooting and Test Functions 3Test FailedProbable Causes• Absorbing solvent or air bubble in flow cel

Strany 306 - Sound Emission

88 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsIntensity TestThe intensity test measures the intensity of the deuteriu

Strany 307 - UV-Radiation

1100 Series DAD and MWD Reference Manual 89Troubleshooting and Test Functions 3Intensity Test EvaluationThe Aligent ChemStation evaluates four spectra

Strany 308 - Solvent Information

1100 Series DAD and MWD Reference Manual 9Source Lens (Achromat) Assembly 199Cell Support Assembly 200Control Module 201Sheet Metal Kit (G1315A/G1365A

Strany 309

90 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsCell TestThe cell test measures the intensity of the deuterium and tung

Strany 310

1100 Series DAD and MWD Reference Manual 91Troubleshooting and Test Functions 3Dark-Current TestThe dark-current test measures the leakage current fro

Strany 311

92 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test FunctionsTest FailedProbable Causes• Defective optical unit.Suggested Actions✔ E

Strany 312

1100 Series DAD and MWD Reference Manual 93Troubleshooting and Test Functions 3Filter TestThe filter test checks the correct operation of the filter a

Strany 313

94 1100 Series DAD and MWD Reference Manual3 Troubleshooting and Test Functions

Strany 314

95Agilent 1100 Series Diode Array and Multiple Wavelength DetectorsReference ManualAgilent Technologies4Repairing the DetectorIntroduction into Repair

Strany 315

96 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorIntroduction into Repairing the DetectorSimple RepairsThe detector is designed for

Strany 316

1100 Series DAD and MWD Reference Manual 97Repairing the Detector 4Cleaning the DetectorThe detector case should be kept clean. Cleaning should be don

Strany 317

98 1100 Series DAD and MWD Reference Manual4 Repairing the DetectorUsing the ESD StrapElectronic boards are sensitive to electronic discharge (ESD). I

Strany 318 - *G1315-90005*

1100 Series DAD and MWD Reference Manual 99Repairing the Detector 4Overview of the Repairing of the DetectorFigure 29 shows the main assemblies and th