2- 60 Agilent B1500 User’s Guide, Edition 7
Introduction
Specifications
Selector
Agilent 16440A SMU/pulse generator selector switches either a SMU or PGU to the
associated output port. You can expand to four channels by adding an additional
16440A. The PGU port on channel 1 provides a “PGU OPEN” function, which can
disconnect the PGU by opening a semiconductor relay. Agilent B1500A and
16445A are required to use the 16440A.
The following specifications data is specified at 23 °C ±
5
°C (
73
°F ±
9
°F) and
50 % relative humidity.
• Channel configuration: 2 channels (CH 1 and CH 2). Can add the additional 2
channels (CH 3 an
d CH 4) by adding another 16440A (selector expander).
• Voltage and current range
Selector Adapter
Agilent 16445A selector adapter is required to control and to supply DC power to
the 16440A selector.
• Power requirement: 100 to 240 V, 50/60 Hz
• Maximum volt-amps (VA): 20 VA
Input Output
Channel 1 (CH 1)
2 (SMU and PGU
a
)
a. PGU channels 1 and 3 have a built-in series semiconductor relay.
1
Channel 2 (CH 2) 2 (SMU and PGU) 1
Channel 3 (CH 3)
b
b. Available when a selector expander is installed.
2 (SMU and PGU
a
)
1
Channel 4 (CH 4)
b
2 (SMU and PGU) 1
Input port Maximum voltage (V) Maximum current (A)
SMU 200 1.0
PGU 40
0.2
a
a. This is peak-to-peak ac current.