Agilent B1500 Users Guide, Edition 7 2- 7
Introduction
Front View
Front View
This section describes the front view of Agilent B1500A.
1. Standby switch
Turns the B1500A on. Pressing the button in the ON state makes the B1500A in
the standby s
tate. The green LED lights when the B1500A is in the ON state.
NOTE When turning the B1500A on
Open the measurement terminals at the device side when turning the B1500A on.
Also disconnect the device from the measurement terminals and open the
measurement terminals after the measurement. If you leave the connection with the
device, the device may be damaged by unexpected operations or charge-up of
measurement cables.
2. HDD access indicator
This green LED lights in the access status of HDD or DVD-ROM/CD-ROM/
CD-RW drive.
Do not turn the B1500A off during this LED lights.