7- 8 Agilent B1500 User’s Guide, Edition 7
Function Details
I/V Sweep Measurement
Synchronous Sweep Measurement
For synchronous sweep measurement, you set up a synchronous sweep source
(VAR1’) in addition to a primary sweep source (VAR1).
The relationship between the output of primary and synchronous sweep sources is
determ
ined by the following equation:
synchronous output = primary outp
ut × ratio + offset
The synchronous output determined by above equation must not exceed the output
range of synchron
ous sweep source.
The following parameter settings are effective for the VAR1’ sweep source.
• Direction (Single or Double)
• Linear/Log
The following setup is required to perform the synchronous sweep measurements.
See “I/V Sweep” on page 5-4 for the GUI.
1. Select the following on the Channel Setup tab screen.
• Select VAR1 for the Function of the SMU used for VAR1.
• Select VAR1’ for the Function of the SMU used for VAR1’.
• Select V, VPULSE, or COMMON for the Mode
of current moni
tor SMU, or
I or IPULSE for voltage monitor SMU. Pulse is available only for one SMU.
Mode of VAR1 and VAR1’ must be the same, voltage or current.
2. Set the VAR1 channel as shown in “Basic Sweep Measurement”.
3. Specify the following parameters for VAR1’ on the Measurement Setup tab
s
c
reen.
Of
fset O
f
fs
et between outputs of primary and synchronous sweep
sources.
Ratio Rat
io between outputs of primary and synchronous sweep
sources.
Compl
iance Compliance value of synchronous sweep source. Allowable
range of
compliance depends on the compliance range of
synchronous sweep source.
Pwr Comp (Optional) Power compliance value of synchronous sweep
source. Allo
wable range of power compliance depends on
the power compliance range of synchronous sweep source.