2- 44 Agilent B1500 User’s Guide, Edition 7
Introduction
Specifications
SMU Common Specifications
Output terminal/connection
Dual triaxial connector, Kelvin (remote sensing)
Voltage/current compliance (limiting)
The SMU can limit output voltage or current to prevent damaging the device
u
nder test.
Voltage compliance:
0 V to ± 100 V (MPSMU, HRSMU)
0 V to ± 200 V (HPSMU)
Current compliance:
± 10 fA to ± 100 mA (HRSMU with ASU)
± 100 fA to ± 100 mA (HRSMU)
± 1 pA to ± 100 mA (MPSMU)
± 1 pA to ± 1 A (HPSMU)
Compliance accuracy:
Same as the current or voltage set accuracy.
About measurement accuracy
RF electromagnetic field and SMU measurement accuracy
Voltage and current measurement accuracy can be affected by RF
electro
magnetic field s
trengths greater than 3 V/m in the frequency range of
80 MHz to 1 GHz. The extent of this effect depends upon how the
instrument is positioned and shielded.
Induced RF field noise and SMU measurement accuracy
Voltage and current measurement accuracy can be affected by induced
RF
field noise strengths greater than 3Vrms in the frequency range of 150 kHz
to 80 MHz. The extent of this effect depends upon how the instrument is
positioned and shielded.
Pulsed measurement
Width: 500 μs to 2 s
Period
: 5 ms to 5 s