Agilent Technologies B1500A Uživatelská příručka Strana 318

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 628
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 317
5- 50 Agilent B1500 User’s Guide, Edition 7
Classic Test Definition
ADC and Integration Time Setup Window
ADC and Integration Time Setup Window
This window is opened by clicking the ADC/Integ button on the I/V Sweep / Multi
Channel I/V Sweep / I/V List Sweep / I/V-t Sampling Measurement Setup screen,
and is used to set the A/D converter of the measurement channels.
A/D Converter This area selects the ADC type and enables
/disa
bles the auto zero function. See
“Integration Time” on page 7-57.
Unit Source monitor unit (SMU) used for the measurement
Name V
ariable name for the measurement data
ADC ADC type, high resolution ADC or high speed ADC
Integration Time This area sets the ADC mode and its factor for each ADC type.
Mode ADC operation mode, auto, manual, or PLC
Factor Factor value. This is the N v
alue show
n in Table 7-8.
Auto Zero For the high resolution ADC. Auto zero function ON or OFF.
See “ADC Zero Function” on page 7-57.
Zobrazit stránku 317
1 2 ... 313 314 315 316 317 318 319 320 321 322 323 ... 627 628

Komentáře k této Příručce

Žádné komentáře