Agilent B1500 Users Guide, Edition 7 4- 35
Using EasyEXPERT
Calibration
Calibration
The Calibration window is opened by clicking the Calibration button, and is used to
perform the self-calibration, SMU current offset cancel, CMU data correction, and
so on.
“Module Self Calibration”
“SMU Zero Cancel”
“CMU Calibration”
Module Self Calibration
The Module Self Calibration tab screen is used to perform the self-calibration of
modules. Check the left check box of Name to select the module for self-calibration.
Open the measurement terminals at the cable end before performing the
self-calibration.
Module Self
Cali
bration
Displays
the following values.
Name Name of module.
Full Range Full range calibration On or Off. Only for the HRSMU/ASU.
Status Calibration status. Pass or fail.
Notes Error information when calibration failed.
Select All Selects all modules for calibration. This checks the check box for all modules.
Unselect All Unselects all modules. This unchecks the check box for all modules.
Start Calibration Starts calibration for the selected modules.
Abort Aborts calibration.
Full Range
Cali
bration
Check
this box to enable the 1 pA range calibration of the HRSMU/ASU.
If you do not use the 1 pA range, uncheck this box. The B1500A does not perform
the 1 pA range offset data meas
urement and calibration.