2- 30 Agilent B1500 User’s Guide, Edition 7
Introduction
Measurement Units
Table 2-14 Measurement Range for Fixed Ranging Mode
Figure 2-6 Impedance vs Frequency Characteristics of Capacitive Load, Calculation
Example
Impe
dance Z
Measurement range (impedance range)
1 kHz ≤ f ≤ 200 kHz 200 kHz < f ≤ 2 MHz 2 M
Hz < f ≤ 5 MHz
0 ≤ Z < 100 Ω 50 Ω 50 Ω 50 Ω
100 Ω ≤ Z <
300 Ω 100 Ω 100 Ω 100 Ω
300 Ω ≤ Z < 1 kΩ 300 Ω 300 Ω 300 Ω
1 kΩ ≤ Z < 3 kΩ 1 kΩ 1 kΩ 1 kΩ
3 kΩ ≤ Z < 10 kΩ 3 kΩ 3 kΩ 3 kΩ
10 kΩ ≤ Z < 30 kΩ 10 kΩ 10 kΩ
30 kΩ ≤ Z < 100 kΩ 30 kΩ 30 kΩ
100 kΩ ≤ Z < 300 kΩ 100 kΩ
300 kΩ ≤ Z 300 kΩ
10 n7
100 n7
1 M7
10 M7
100 M7
1 m7
10 m7
100 m7
1 7
10 7
100 7
1 k7
10 k7
100 k7
1 M7
10 M7
100 M7
1 G7
10 G7
100 G7
1 T7
10 T7
1 kHz 10 kHz 100 kHz 1 MHz 5 MHz
10 MF
100 MF
1 mF
10 mF
100 mF
1 F
10 pF
100 pF
1 nF
10 nF
100 nF
1 MF
100 fF
1 pF
10 fF
100 aF
1 fF
300 k7 range
50 7 range
3 k7
range
Komentáře k této Příručce