
2- 48 Agilent B1500 User’s Guide, Edition 7
Introduction
Specifications
MFCMU
MFCMU: Multi Frequency Capacitance Measurement Unit
• Measurement functions
• Measurement parameters: Cp-G, Cp-D, Cp-Q, Cp-Rp, Cs-Rs, Cs-D, Cs-Q,
Lp-G, Lp-D, Lp-Q, Lp-R
p, Ls-Rs, Ls-D, Ls-Q, R-X, G-B, Z-θ, Y-θ
• Ranging: Auto or fixed
• Measurement terminal: Four-terminal pair configuratio
n, four BNC (female)
connectors
• Cable length: 1.5 m or 3 m, automatic identification of accessories
• Test signal
• Frequency
Range: 1 kHz to 5 MHz
Resolution: 1 mHz (minimum)
Accuracy: ± 0.008 %
• Output signal level
Range: 10 mVrms to 250 mVrms
Resolution: 1 mVrms
Accuracy:
± (10.0 % + 1 mVrms) at the measurement port of the MFCMU
± (15.0 % + 1 mVrms) at the measurement port of the MFCMU cable (1.5 m
o
r 3.0 m)
• Ou
tput impedance: 50 Ω, typical
• Signa
l level monitor
Range: 10 mVrms to 250 mVrms
Accuracy (open load):
± (10.0 % of reading + 1 mVrms) at the measurement port of the MFCMU
± (15.0 % of reading + 1 mVrms) at the measurement port of the MFCMU
cab
le (1.5 m o
r 3.0 m)
Komentáře k této Příručce