10-76 Agilent B1500 User’s Guide, Edition 7
Application Library and Utilities
Desktop EasyEXPERT
Advanced Setup Differences on Advanced Setup
Series R always shows NONE.
Wait Time Control is not supplied.
Restrictions for
Standard Edition
There are the following restrictions for the Desktop EasyEXPERT Standard edition.
I/V-t Sampling classic test is not supported.
VSU, VMU, and PGU are not supported.
Agilent E5250A Switching Matrix is not supported.
Table 10-17 Setting of High Resolution ADC
NOTE To reuse the test setup for B1500A
Perform the following procedure to reuse the test setup created for the B1500A.
1. Launch Desktop EasyEXPERT with the condition Set Analyzer=B1500A.
2. Export the test setups to be reused. And exit Desktop EasyEXPERT.
3. Launch Desktop EasyEXPERT with the condition Set Analyzer=4155B, 4156B,
4155C, or 4156C.
4. Import the test setups exported at the step 2.
5. Set the channel for the analyzer to use.
Also, the procedure made by replacing the step 1 with the step 3 can be used to reuse
the test setup created for the 4155/4156.
Mode Factor Setting of 4155/4156
AUTO - MEDIUM
MANUAL 1 N 127 SHORT, TIME=80 μs × N
PLC 1 MEDIUM
N 2LONG, NPLC=N