104 8000 Operator’s Manual
4 Operating the 8000
Following is a description of the Print Selections screen options:
Table 2 7 Print Selections Options
Option Function
1 Batch Information Select option 1, Batch Information, to immediately print the previously entered
batch information, if any, for the selected program. This printed data is the header
information you can enter via option 4, Headers, on the Program Variables screen.
2 Sample Times Select option 2, Sample Times, to immediately print the sampling times for each
row of the selected program. This is the complete listing of all time points entered
in option 2, Sample Time Points, under the Program Variables screen.
3 Turn On (Off) Rep Center Select option 3, Turn On (Off) Rep Center, to enable or disable the Report Center
Printer. This option always reflects the opposite of the current state of the Report
Center Printer. For example, if it is currently disabled,
TURN ON REP CENTER
displays next to option 3.
Note: The state of the Report Center Printer is not unique to an individual program.
It is global. When you enable or disable it, the Report Center Printer stays enabled
or disabled until you change it again, even if you switch programs.
4 Set Com Port ID Select option 4, Set Com Port ID, to enter a new communication port identification
number. Your system was shipped with the communication port identification
number set to 01. You can set the identification number to any number between 1
and 99. The identification number is used when multiple units are connected in
series so that the controlling or master 8000 can start individual units.
Enter the communication port identification number and press ENTER. The Print
Selections screen displays.
5 Set Baud Rate Select option 5, Set Baud Rate, to set the baud rate for your system. If you are using
the 8000 with other Agilent equipment, set the baud rate to 9600 baud (option 4).
Select the desired baud rate from the menu. The Print Selections screen displays.
Note: In order to set the baud rate from this option, the baud rate must be set to
programmable baud (see MENU > 0 > 8 under “Hidden Key Functions” on
page 41.