36 8000 Operator’s Manual
3 Setting Up the 8000
Installing the Sample Tray
Place the supplied sample tray on top of the unit. The sample tray has
four grommeted corner holes which fit over the four locating pins.
Installing the Media Rinse Reservoir
The media rinse reservoir fits in the rear of the unit between the metal
brackets. The dispensing arm must be moved forward to put it in place.
1 From the Ready screen, press MENU.
2 Select option 7, Manual Operation.
3 Select option 4, Goto Rows.
4 Enter 4 to move the dispensing arm to Row 4. The dispensing arm
moves forward to allow access to the rear of the 8000.
5 Place the media rinse reservoir between the metal brackets directly
behind the sample tray.
6 Connect the six-foot length of clear Tygon tubing to the barbed
elbow fitting located on the rinse chamber of the reservoir. Place the
other end in a sink, drain, or suitable waste container.
7 If your system configuration includes the syringe pump, press H to
move the dispensing arm to the home position and press ESC three
times to return to the Ready screen.
The sample tray must be securely mounted on the locating pins.
Otherwise, damage can occur to the sampling needles and drive
mechanism.
The media rinse reservoir has two chambers—one for replacement
media and one for rinsing. The replacement media chamber has two
barbed fittings—one on each end of the tank. The rinse chamber has a
single barbed fitting. Orient the reservoir with the rinse chamber toward
the front.