Operating the 8000 4
8000 Operator’s Manual 47
The front panel options include the following:
Table 2 Front Panel Options (Apparatus 1 / 2)
Option Function
ESC Press ESC to stop execution of a running program or calibration routine. Press
ESC to move back one level to the previous menu or selection when moving
through the multi-level menu.
Note: Be very careful when using ESC while a program is running. If a program
or calibration is running and ESC is pressed, a prompt appears on the first line of
the display asking if you want to stop the running program. The options are Y or
N. Press Y to stop the running program. Press N to continue the running
program.
MENU Use this option to set the clock, alarms, DISSO parameters, start output delay /
syringe pump parameters, manual prime time / volume, manual purge time /
volume as well as to remotely control the dissolution apparatus and manually
control the 8000.
PRINT Press PRINT to record the batch information, print the sample times, enable or
disable the Report Center Printer, select a local or remote printer, set the
communication port identification number, and set the baud rate.
CAL Use this option to calibrate the 8000 to ensure that sample volumes are
accurately and precisely taken.
Note: The CAL key on the 8000 can be used only if the peristaltic pump is
installed on your system.
PROG Use this option to set program parameters.
START PROG Use this option to start a program.
MANUAL SAMPLE Use this option to collect a sample from the dissolution vessels on command
without a program.
OPEN VALVES Press OPEN VALVES to open all valves simultaneously. This function keeps
the valves open as long as the key on the keypad is pressed. To change the valve
configuration, press MENU > 0 > 0 and select the appropriate configuration.
(PROBES UP / DOWN)
Press (PROBES UP / DOWN) to raise or lower the sampling cannulas on
the dissolution apparatus.
Note: Not available with APP 3/7 systems.