Agilent Technologies 8000 Series Návod k obsluze Strana 79

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 126
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 78
Operating the 8000 4
8000 Operator’s Manual 79
Main Menu Option 2 - Set APP 3/7
From the Main Menu, select option 2, Set APP 3/7. Enter a program
number and press ENTER. The Set APP 3/7 Parameters screen
displays.
Following is a description of the Set APP 3/7 Parameters screen options:
***SET APP 3/7 PARAMETERS***
1 DIPPING SPEED (DPM) 2 SAMPLES PER ROW
3 HOLD DIP TIME 4 SET BATH TEMP
5 COM PORT ID 6 PRINT FREQUENCY
Table 1 7 Set App 3/7 Parameters Options (Apparatus 3 / 7)
Option Function
1 Dipping Speed (DPM) Select option 1, Dipping Speed (DPM). Enter the desired DPM and press
ENTER. The Set APP 3/7 Parameters screen displays.
2 Samples Per Row Select option 2, Samples Per Row. Enter the desired number of samples to be
taken from each row and press ENTER. The Set APP 3/7 Parameters screen
displays.
3 Hold Dip Time Select option 3, Hold Dip Time. Enter a hold dip time and press ENTER to
advance to each remaining row. After a hold dip time is entered for the final row,
the Set APP 3/7 Parameters screen displays.
4 Set Bath Temp Select option 4, Set Bath Temp. Enter the desired water bath temperature and
press ENTER. The Set APP 3/7 Parameters screen displays.
Zobrazit stránku 78
1 2 ... 74 75 76 77 78 79 80 81 82 83 84 ... 125 126

Komentáře k této Příručce

Žádné komentáře