Operating the 8000 4
8000 Operator’s Manual 85
Following is a description of the Operate Syringe/Filter Changer screen
options:
Table 2 1 Operate Syringe/Filter Changer Options
Option Function
1 Valve to Input The syringe pump contains a valve to control the direction of the flow of medium.
Use this option to move the valve to the input position to manually input or fill the
syringe with medium.
Select option 1, Valve to Input, to move the valve to the input position. The valve
emits a loud click when it moves. If you hear no sound, the valve is already in the
input position.
2 Valve to Output The syringe pump contains a valve to control the direction of the flow of medium.
Use this option to move the valve to the output position to manually output or
purge the syringe of medium.
Select option 2, Valve to Output, to move the valve to the output position. The
valve emits a loud click when it moves. If you hear no sound, the valve is already
in the output position.
3 Syringe Fill Use this option to draw a specific amount of medium into the syringes. Ensure the
valve is in the input position before drawing the medium (see “1 Valve to Input”
above).
Select option 3, Syringe Fill. Enter the desired percentage of each syringe to fill
and press ENTER. The syringes draw the specified amount of medium.
4 Syringe Dispense Use this option to dispense medium from the syringes.
Select option 4, Syringe Dispense. The syringes dispense the previously drawn
medium (see “3 Syringe Fill” above).
Zobrazit stránku 84