22 8000 Operator’s Manual
3 Setting Up the 8000
Unpacking / Setup of 8000 & Peristaltic Pump
Environmental Requirements for Peristaltic Pump
Installation
Altitude: max 2000m;
Humidity: max relative humidity 80% for temperatures up to 31 ºC
decreasing linearly to 50% relative humidity at 40 ºC
Indoor use only
Temperature: 5 ºC to 40 ºC
Main supply voltage fluctuations are not to exceed ± 10% of the
nominal supply voltage
Operating conditions: 3 mins ON Max; 5 mins OFF Min
Figure 2 Peristaltic Pump
Installation
1 Carefully remove the peristaltic pump from its shipping carton.
Remove all cushioning and tape.
2 Place the peristaltic pump on a clear, dry, level section of the bench
top between the 8000 and the dissolution apparatus.
3 Remove the two screws on top of the pump. Lift off the plastic cover
and set it aside.
4 Remove the cartridges from the pump by pressing down the levers on
the right side. Lift the cartridges up and away from the pump.