118 8000 Operator’s Manual
5 Maintenance and Troubleshooting
9 Press ESC to return to the Manual Operations screen.
10 Select option 5, Turn On R-M Pump.
11 Allow the pump to run until the source beaker is empty.
12 Select option 5, Turn Off R-M Pump.
13 Repeat Steps 2 - 12 using ultrapure water in place of the warm water.
14 Select option 5, Turn On R-M Pump.
15 Purge the system by removing the source line from the ultrapure
water and allowing the pump to run until water no longer flows
through the return line.
16 Select option 5, Turn Off R-M Pump.
17 From the Manual Operations screen, select option 4, Goto Rows.
18 Press 6 to move the dispensing arm to Row 6.
19 Remove the rinse reservoir, empty any fluid, and replace the
reservoir.
20 Press H to return the dispensing arm to the home position.
21 Press ESC until the Ready screen displays.
If additional cleaning is required, repeat the above procedure replacing
the warm water with diluted alcohol (1:10).