32 8000 Operator’s Manual
3 Setting Up the 8000
between the remaining extension of the Y cable attached to the
syringe pump and the RS232 port on the dissolution apparatus.
10 A system configuration that includes a non-Agilent dissolution
apparatus, requires use of an adapter (P/N 5075-0852, see following
Figure). This adapter should be attached to the #2 port of the Y-cable
in place of the cable connected to the dissolution apparatus.
Figure 10 Y Cable 5075-0446 and Adapter 5075-0852
11 The 8000 is shipped with all sample tubing attached to the valves
inside the dispensing arm.
12 If the system configuration includes the 705-DS, connect the resident
probes to the sample tubing and attach the lines to the left port
(labeled IN) of each syringe starting with Position 1.
Otherwise, connect the tubing from the dissolution apparatus
sampling cannulas to the left port (labeled IN) of each syringe
NOTE
Do not remove the nylon tie wraps securing the tubing into bundles.