94 8000 Operator’s Manual
4 Operating the 8000
Table 2 4 App 3/7 Options
Option Start Now Function Delayed Start Function
1 Load &
Run APP 3/7
Select option 1, Load & Run APP
3/7, to upload all saved parameters
to the Apparatus 3 / 7 and start the
program.
Select option 1, Load & Run APP 3/7, to upload all saved
parameters to the Apparatus 3 / 7.
Enter the start time in 24-hour format and press ENTER. Enter the
start date and press ENTER.
If the delayed start alarm is enabled, enter the duration before the
start of the program you would like the alarm to sound. To disable
the alarm, see MENU > 0 > C under “Hidden Key Functions” on
page 41.
At any time prior to the scheduled start of the program, press ESC
to cancel the delayed start.
2 Run WO/
Loading
Select option 2, Run WO/
Loading, to start the program and
run the 8000 without loading and
overriding the Apparatus 3 / 7
settings.
Select option 2, Run WO/ Loading, to run the 8000 without loading
and overriding the Apparatus 3 / 7 settings.
Enter the start time in 24-hour format and press ENTER. Enter the
start date and press ENTER.
If the delayed start alarm is enabled, enter the duration before the
start of the program you would like the alarm to sound. To disable
the alarm, see MENU > 0 > C under “Hidden Key Functions” on
page 41.
At any time prior to the scheduled start of the program, press ESC
to cancel the delayed start.
3 Run 8000
Only
Select option 3, Run 8000 Only, to
immediately start the 8000 only.
The 8000 executes the program
currently selected.
This option is used when other
instrumentation, such as an
Apparatus 3 / 7, is connected to
the 8000 through the START
OUTPUT jack on the rear of the
instrument.
Select option 3, Run 8000 Only, to start the 8000 only.
Enter the start time in 24-hour format and press ENTER. Enter the
start date and press ENTER. Once the start date and time are
reached, the 8000 executes the program currently selected.
If the delayed start alarm is enabled, enter the duration before the
start of the program you would like the alarm to sound. To disable
the alarm, see MENU > 0 > C under “Hidden Key Functions” on
page 41.
At any time prior to the scheduled start of the program, press ESC
to cancel the delayed start.
4 Set APP
3/7
Use this option to modify or review the current program parameters. Select option 4
, Set APP 3/7. The
Set APP 3/7 Parameters screen displays. See “Main Menu Option 2 - Set APP 3/7” on page 79 for a
description of the Set APP 3/7 Parameters screen options.